C96IB01
57
EN
2.
Empty the measuring cell by pressing the pump lever.
3.
Open the pump lock and remove the pump.
Fig. 11 - 1
Removing the pump
4.
Screw out the screw of the pump lock using an Allen key size 2.5 and remove
the screw.
Fig. 11 - 2
Removing the pump lock screw
5.
Remove the pump lock by sliding it outwards.