![Anthogyr Implanteo Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/anthogyr/implanteo/implanteo_instruction-manual_2947759018.webp)
46
10.1 Device maintenance
10. Cleaning, disinfection and sterilisation
Single use
Non sterilisable
Sterilisable in an autoclave
at 134°C for 18 minutes
Micromotor sterilisation cap
Sterilisable
non immersible
(= disinfection)
Sterilisable
and immersible
(+ ultrasound)
w
The bags of physiological
fl uid
w
The irrigation lines
w
Screen protection
w
Micromotor protective
sheath
w
Control unit
w
Pedal
w
Calibration module
w
Micr cable
w
Bracket
w
Motor support
w
Mandrel
w
Fixing clips for
the irrigation line
At the end of the dental surgery :
w
Put the supply switch in the O position.
w
Remove the tool from the contra-angle.
w
Remove the bottle or bag from the bracket.
w
Remove the perforator of the irrigation line from the bottle or bag of physiological serum or sterile water.
w
Remove the clips for irrigation lines.
w
Disconnect the contra-angle from the micromotor.
9. Stopping the device
If the last protocol sequence is active,
a short press on button « C »
makes the fi rst sequence of this protocol active.
The fi rst protocol sequence is active,
a long press on button « C »
makes the last sequence of this protocol active.
C
C
8. Using the pedal
The lever
gradually adjusts the speed according
to the motor during the operational phase.
A short press on the button « A »
is a 20% increase in
the speed of the fl ow of the peristaltic pump.
A long press on the button « A »
starts the pump at
the programmed fl ow , without using the micromotor,
or the tool.
A short press on the button « B »
inverts the rotation
direction of the tool.
A long press on the button « B »
increases
the limitation of the torque in increments of 5 N.cm.
A short press on the button « C »
goes on to the next
sequence for the active protocol.
A long press on the button « C »
goes to the previous
sequence for the active protocol.
Button A
Pump
Lever
Speed
Button B
Rotation direction
of micromotor
/ Increase in
torque
Button C
Changing
sequence
*All of this information is summarised in a maintenance poster supplied with the motor.