![Anthogyr Implanteo Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/anthogyr/implanteo/implanteo_instruction-manual_2947759014.webp)
42
Data « torques and speeds » are located in the index “ImplMotCurve“.
A fi le called « name » is created during backup. « name » is corresponding to what you captured.
A fi le called « name.csv » is created there during backup, if the fi le already exists, an incremental fi le is added : “Name001.csv“.
It is a text fi le format that can be imported into a spreadsheet (Microsoft Excel for example).
Those fi les can be directly imported into a spreadsheet.
The following data are measured every 0.25 s :
TIME :
display time (in ms) that passed since the device is switched on.
TORQUE :
display torque in 1/100 of N.cm (or in 1/100 of % if the reduction ratio is different from 20:1).
SPEED :
display motor speed, in rpm.
.png images format
(Fig.8)
are also created in the index « name », one fi le for each sequence where recording
of torque and speed
(Fig.5, Rep.5)
is activated.
6.5 Saving torque and speed data
The procedure to follow for saving « torque and speed data » is as follows :
Press the button
(Fig.3, rep.8)
.
Press the button
(Fig.6, rep.7)
.
Name the fi le and then validate.
Fig. 8
Fig. 7
6.6 Changing a contra-angle
This is the list of available contra-angles :
30:1
20:1
16:1
5:1
1:1
1:3
256:1
Press the icon for the reduction ratio for the active contra-angle
(Fig.3, rep.2).
Note that the speed ranges at the exit of the tool are displayed to the side of the selected contra-angle.
C
20:1