CHAPTER 8
69
C H A P T E R 8
Specifications DECADE Elite
Environmental, dimensions, weight & power requirements
Working temperature
Storage temperature
10 - 40°C (indoor use only)
–25 - +50°C
Humidity
20 - 80% RH
Safety and EMC
According to EC-directives; Emission Group I
Class A; cMETus approved
Equipment class
1
Installation category
II
Pollution degree
2
Dimensions
43 (D) x 22 (W) x 44 (H) cm = 16.9” (D) x 8.7”
(W) x 17.3” (H)
Weight
14.4 kg (32 lbs) without flow cell and column
Installation
Install upright on flat & smooth surface, keep
space under the detector free (risk of blocking
power supply fan).
Power requirements
110-240 VAC, 50/60 Hz, 260 VA, auto-sens-
ing
Mains fuse
2.5 AT / 250V, 5x20 mm, IEC 60127-2
For safety reasons, any other internal fuse or
circuit breaker is not operator accessible, and
should be replaced only by Antec authorized
personnel. Use only manufacturer-supplied
fuses.
For optimum analytical performance it is recommended that the
ambient temperature of the laboratory be between 17-25 °C and be held
constant to within ± 2 °C throughout the entire working day. For optimal
performance of the oven the temperature should be set at least 7
degrees above ambient.