Chapter 4
DECADE Elite FW upgrade
31
•
Click “Yes” when the unit is fully started, only then it will automati-
cally re-
connect. In case of any problem, click “No”, switch off and
on the unit and repeat this step (click button 5).
•
After clicking “Yes” the unit will automatically connect, and read
the device parameters. This takes a few seconds, and is followed
by opening the settings file (created in step 1.). The calibration set-
tings will be send to the instrument
after clicking “OK”.
•
Close Elite Dialogue.
•
Switch the detector off and on again.
•
Start the Elite Dialogue software and reconnect the instrument.
•
Goto ‘Options’ menu, Device settings or click the ‘F12’ button.
•
Check if the serial numbers (left side of the device settings win-
dow) are filled in and correct.
•
If correct the FW upload procedure was successfully completed.
Switch cables and connect the unit with a LAN cable to the PC in case
an OQ, PQ or any other measurement is done using Elite Dialogue.
Switch off and on the unit and reconnect using the predefined LAN port
(192.168.5.1).