Timing Menu
Input
SERVICE MENU:
TIMING
ALARM DELAY
30 SEC
This is the time between an alarm being sensed and the alarm sounding (5-3,600sec). This gives the
sensor time for the pressure to correct itself without alarm being sounded.
Default 30 seconds
CAUTION IN DELAY
NO
Whether a caution is displayed during the alarm delay time.
Default: No
RPI MUTE TIME
15 MIN
The time between mute button being pressed and an alarm sounding once more (1-1440 minutes)
Default: 15 minutes
PRESS MENU
TO EXIT
PRIMARY SENSOR
REVERSE
Room Mode (default) will display Setback/Isolation. This mode can also be
displayed as
Occupied/Unoccupied.
NOTE:
This variable cannot be change if a Keyswitch nurse station is con-
nected.
SENSOR SCALE
FACTOR
This allows the primary sensor reading to be ajusted to match a manual
reading taking during setup/balancing. It provides a linear scaling of the
sensor reading and is the preferred method of adjustment. (range: 0.500
to
2.000)
Default: 1.000
SENSOR OFFSET
This allows for the pressure reading to be offset by a fixed amount. This
can be used in conjunction with the sensor scale to provide a 2-point linear
calibration (y=mx+b).
In most cases this is not required and should be left at 0.000. Sensor scale is
the preferred method of adjusting the reading.
(Range: -0.0500
to
0.0500
in.w.c.
)
Default: 0.0000
in.w.c.
PRESS MENU
TO EXIT