Error Messages
Self Test Error Messages
A listing of Self Test Error messages is provided in Table 2-2.
Chapter 2 Functions and Operations
2-47
Error Message
Description
Battery Low
Battery voltage is less than 9.5 volts. Charge battery.
If condition per-
sists, call your Anritsu Service Center.
External Power Low
External supply voltage is less than 10 volts.
Call your Anritsu Service
Center
PLL Failed
Phase-locked loops failed to lock. Charge battery.
If condition persists
with a fully charged battery, call your Anritsu Service Center
Integrator Failed
Integration circuit could not charge to a valid level. Charge battery.
If
condition persists with a fully charged battery, call your Anritsu Ser-
vice Center
.
EEPROM R/W
Failed
Non-volatile memory system has failed.
Call your Anritsu Service
Center.
Out Of Temp.
Range
Ambient temperature is not within the specified operating range.
If the
temperature is within the specified operating range and the condition
persists, call your Anritsu Service Center.
RTC Battery Low
The internal real-time clock battery is low. A low or drained clock bat-
tery will affect the date stamp on saved traces.
Contact your nearest
Anritsu Service Center.
LO Lock Fail
The local oscillator in the spectrum analyzer has phase lock loop er-
rors.
If condition persists with a fully charged battery, call your Anritsu
Service Center
.
Battery Cal Lost
Battery communication failed. The indicated battery charge status may
be invalid.
If condition persists, call your Anritsu Service Center.
Memory Fail
The EEPROM test on the Site Master main board has failed.
If condi-
tion persists, call your Anritsu Service Center.
The time and date
Have not been set
on this Site Master.
To set it, after exit-
ing here press
the <SYS> [Clock]
keys.
Press ENTER or
ESC to continue
The time and date are not properly set in the Site Master.
If condition
persists, call your Anritsu Service Center.
Note:
A listing of Anritsu Service Centers is provided on page 1-11.
Table 2-2.
Self Test Error Messages