IP Network Measurement Division
27
40G Quick Start Guide
(
Jitter
)
MP1803A MUX Setting for Data output
MP1803A MUX Setting for Data output
1
2
3
4
5
In case of MP1803A with option 01
1.
Turn “Data/Data Tracking” to ON.
2.
Set “Data Amplitude to 1.000 V.
3.
Set “Data Cross Point” to 50.0%.
4.
Set “Offset” to Vth.
5.
Set “Data Threshold” to 0.000 V.
In case of MP1803A without option 01
No setting for Data output.
Data output is fixed 2.0Vp-p (Add 6dB)
IP Network Measurement Division
28
40G Quick Start Guide
(
Jitter
)
MP1804A DEMUX Setting for Data input
MP1804A DEMUX Setting for Data input
1
1.
In case connect to MP1797A,
Set “Data Threshold” to 0.000V (- 0.130 V typ. In case 6dB added, at 43G)
Need to adjust optimum value.
2.
Set “1 /4 Output Clock Delay” to 0.0 ps (Fixed).
2