IP Network Measurement Division
25
40G Quick Start Guide
(
Jitter
)
Activation
Activation
Starting and Terminating the System
Starting procedure
1. Before connecting MP1803A/MP1804A to PPG, SG(MP1797A) and ED,
turn ON PPG and turn off output.
2. Connect MP1803A to PPG, SG and MP1804A to ED.
3. Turn ON ED.
4. Turn ON the main power switch on the rear panel of the MP1804A.
* Turn ON the DUT (When connect the DUT).
5. Turn ON the power switch on the front panel of the MP1803A.
The lamp on the front panel of the MP1803A is turned on.
6. Turn ON the SG(MP1797A).
7. Turn ON the Data output of PPG.
Termination procedure
1. Turn OFF the power for PPG.
2. Turn OFF the power for SG.
3. Turn OFF the main power switch on the rear panel of the MP1803A.
* Turn OFF the DUT (When connect the DUT).
4. Turn OFF the main power switch on the rear panel of the MP1804A.
5. Turn OFF ED.
1. PPG On,
& Output Off
2. Connect
Coaxial cables
13
CAUTION
If the MP1803A and its peripheral equipment are started or terminated without going through
the above steps, the expensive semiconductors built in the MP1803A could be damaged.
3. ED
4. MP1804A
(DUT)
5. MP1803A
6. SG
(Output)
7. PPG
(Output)
S
tar
ting
Pro
ced
u
re
Termin
ation
Pr
ocedu
re
IP Network Measurement Division
26
40G Quick Start Guide
(
Jitter
)
Activation
Activation
Operating the MP1803A
1. Initially, set to Off the DATA outputs of the PPG.
2.
Set combination Mode of MP1800A 4ch PPG.
3. Set DATA AMPLITUDE to 1.0 Vpp and OFFSET voltage VOH to 0 V for all the 4 channels of PPG.
4. Turn on SG(MP1797A).
5. Set Bit rate 39.8Gbps or 43.0Gbps and set Tx Jitter off.
6. Turn on PPG output.
7. Turn the 1/4 Output Clock Delay dial of the MP1803A to generate the optimum 40G waveform.
8. Recommendation:
Observing the waveform using the sampling oscilloscope, turn the 1/4 Clock Delay dial of the MP1803A to find a
point where the best waveform is obtained with the minimum jitter. The delay can be varied in the range of –70 to +70 ps.
How to make Error Free Condition of MP1800A 4ch ED.
1. Set combination Mode of MP1800A 4ch ED.
2. Select the Single-ended interface of DATA input
3. Start Auto adjust Function of MP1800A 4ch ED.
4. Adjust Threshold Adjust Dial for 1/1 DATA Input of MP1804A. (Search error free point)
5. Adjust Phase Adjust Dial for 1/1 CLOCK Input of MP1804A. (Search error free point)