Chapter 2 Specifications
2.1 Specifications
No. Item
Specifications
1 Measurement
accuracy
±
20 mUIpp (with respect to the amount of jitter from the trans-
mitter that conforms to Phase Analysis Method recommended in
ITU-T O.172 (April,2005) Appendix
Ⅷ
.)
1.1 Measurement
conditions
Measurement time: 1 min.
Average: 5
measurements
Filter:
10 G: 20 kHz to 80 MHz, 50 kHz to 80 MHz
2.5 G: 5 kHz to 20 MHz, 12 kHz to 20 MHz
Transmitter:
MU150121A/B, MU150134A, or an optical
transmitter specified by Anritsu
Pattern:
OC-192c/STM-64c (J0/Z0 = “AA” + “CC”)
OC-48c/STM-16c (J0/Z0 = “AA” + “CC”)
Payload:
All 0 (Scrambled to PRBS7)
Optical input power:
−
10 to
−
12 dBm
1.2 Measurement
system
For 10 G
Connection on the left side of MP1590B
(for 10GTx MU150121A/B)
2-2