2-37 Application Options Menu
Spectrum Analyzer
2-102
PN: 10580-00349 Rev. M
Spectrum Analyzer MG
2-37 Application Options Menu
Key Sequence:
Shift
>
System
(
8
) key
> Application Options
2-38 Other Menus
Preset
,
File
,
Mode
, and
System
menus are described in the instrument User Guide.
Impedance 50 ohm 75 ohm Other:
Select either 50 ohm, 75 ohm, or
Other impedance value. Selecting 75 ohm selects the 7.5 dB loss of the
Anritsu 12N50-75B adapter. For other adapters, select Other and enter the
appropriate loss (0 dB is the default loss). Note that loss values are also
applied to the relevant units.
Bias Tee:
Chapter 8, “Bias Tee (Option 10)”
. See the product
Technical Data Sheet to check option availability on your instrument model.
Auto Ref Level:
The Auto Ref Level function adjusts the position of a
signal on the display screen so that it is approximately two divisions down
from the top, if possible. When the key is pressed, the reference level is
adjusted once. Auto Ref Level may turn off the low-noise front-end
preamplifier, but does not turn it on. It has no effect on vertical scaling.
Back:
Press this submenu key to return to the previous menu.
Figure 2-90.
SPA Application Options Menu
Options
Impedance
50 Ohm 75 Ohm Other
Back
Bias Tee
Auto Ref Level
Содержание BTS Master
Страница 103: ...Spectrum Analyzer 2 35 Trace Menu Spectrum Analyzer MG PN 10580 00349 Rev M 2 91 Figure 2 78 Trace Info...
Страница 246: ...6 6 Coverage Mapping Menu Coverage Mapping Option 431 6 24 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 278: ...7 21 Marker Menu AM FM PM Analyzer Option 509 7 32 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 282: ...8 2 Bias Tee Menu Bias Tee Option 10 8 4 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 295: ......