3-12 Measurements Menu
W-CDMA/HSPA+ Signal Analyzer
3-36
PN: 10580-00234 Rev. V
3GPP MG
HSDPA Menu
Key Sequence:
Measurements
> Demodulator > HSDPA
Selected Code:
Press this submenu key to select the active code, using the
arrow keys, rotary knob, or numeric keypad. The range is 0 to 255 or 0 to 511
depending on the Max Spreading Factor set under the Setup main menu.
Refer to
.
Next Active Code Left:
Select the next active code to the left of the selected
code.
Next Active Code Right:
Select the next active code to the right of the
selected code.
Total Time:
Set the time for the power versus time screen. The maximum
total time is 72 hours.
Single Sweep Time:
Set the single sweep time. The instrument
automatically calculates the total time.
IQ Persistence:
Set the number of samples before displaying the screen
(maximum 48).
Control Channels:
Select a control channel mode between Rel, Abs, and
Delta.
Back:
Returns to the
“Demodulator Menu” on page 3-34
Figure 3-27.
HSPA+ Menu
HSDPA
IQ Persistence
1
Back
Next Active
Code
Left
Total Time
240 s
Single Sweep Time
5 s
Control Channels
Rel Abs Delta
Next Active
Code
Right
Selected Code #
2
Содержание BTS Master MT8221B
Страница 1: ...BTS Master User Guide MT8221B and MT8222B The High Performance Handheld Base Station Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 24: ...1 12 Secure Environment Workplace General Information 1 10 PN 10580 00207 Rev E BTS Master UG ...
Страница 46: ...2 13 Firmware Overview Instrument Overview 2 22 PN 10580 00207 Rev E BTS Master UG ...
Страница 54: ...3 8 PN 10580 00207 Rev E BTS Master UG ...
Страница 88: ...7 8 PN 10580 00207 Rev E BTS Master UG ...
Страница 94: ...8 5 Updating Instrument Firmware Master Software Tools 8 6 PN 10580 00207 Rev E BTS Master UG ...
Страница 95: ...Master Software Tools 8 5 Updating Instrument Firmware BTS Master UG PN 10580 00207 Rev E 8 7 ...
Страница 96: ...8 5 Updating Instrument Firmware Master Software Tools 8 8 PN 10580 00207 Rev E BTS Master UG ...
Страница 97: ...Master Software Tools 8 5 Updating Instrument Firmware BTS Master UG PN 10580 00207 Rev E 8 9 ...
Страница 98: ...8 5 Updating Instrument Firmware Master Software Tools 8 10 PN 10580 00207 Rev E BTS Master UG ...
Страница 114: ...C 8 PN 10580 00207 Rev E BTS Master UG ...
Страница 118: ...Index 4 PN 10580 00207 Rev E BTS Master UG ...
Страница 119: ...Anritsu Company 490 Jarvis Drive Morgan Hill CA 95037 2809 P N 10000 00000 Revision Prelim Printed March 2014 ...
Страница 152: ...2 18 Other Menus GSM GPRS EDGE Signal Analyzer 2 18 PN 10580 00234 Rev V 3GPP MG ...
Страница 196: ...3 18 Other Menus W CDMA HSPA Signal Analyzer 3 44 PN 10580 00234 Rev V 3GPP MG ...
Страница 224: ...4 16 Other Menus TD SCDMA HSPA Signal Analyzer 4 28 PN 10580 00234 Rev V 3GPP MG ...
Страница 272: ...5 19 Other Menus LTE Signal Analyzer 5 48 PN 10580 00234 Rev V 3GPP MG ...
Страница 338: ...7 5 Coverage Mapping Menu LTE and TD LTE Coverage Mapping 7 16 PN 10580 00234 Rev V 3GPP MG ...
Страница 354: ...8 4 W CDMA EMF Menu EMF Option 444 8 16 PN 10580 00234 Rev V 3GPP MG ...
Страница 360: ...Index 4 PN 10580 00234 Rev V 3GPP MG ...
Страница 361: ......