Anritsu 3656C W1 Скачать руководство пользователя страница 40

2-9 Application Interface – Main Menu Tab

Using the PVS Software with the ME7838A/AX/E/EX

2-28

PN: 10410-00784  Rev. A

3656C W1 Cal/Ver Kit & 2300-616-R PVS UG

Repeat Last Cal

This command button re-calibrates the VNA according to the CURRENT calibration. This provides a way to 
recalibrate the VNA without having to reset the program parameters in case the original calibration was 
flawed for some reason. When the calibration sequence is complete, program execution continues to the

 Main 

Menu tab

 shown previously in 

Figure 2-36, “ME7838A Application Interface – Main Menu Tab” on page 2-25

.

Save Data

This command button writes data files to the PC Controller and stores the files at 

C:\mmdcsvc\Data

. The 

program does not write data to the verification kit USB memory device. If selected, program execution 
continues as described in 

Section 2-13 “Saving Verification Data” on page 2-36

.

The number of saved files varies depending on the user settings:

If both the matched and mismatched thru tests are run, two files are written to the PC Controller hard 

drive.

If only one of the tests was run, only one file is written to the PC hard drive at 

C:\mmcsvc\Data

Matched Thru

This command button runs only the Matched Thru tests. If selected, program execution continues as described 
in 

Section 2-11 “Matched Thru Tests” on page 2-30

. When the tests are completed, the program returns to the

 

Main Menu tab

 shown in 

Section 2-9 “Application Interface – Main Menu Tab” on page 2-25

.

Matched Thru SN

This field is used to enter the serial number of the matched thru.

Mismatched Thru

This command button runs only the Mismatched Thru tests. If selected, program execution continues in 

Section 2-12 “Mismatched Thru Tests” on page 2-33

When the tests are completed, the program returns to the

 

Main Menu tab

 shown in 

Figure 2-36, “ME7838A Application Interface – Main Menu Tab” on page 2-25

.

Mismatched Thru SN

This field is used to enter the serial number of the matched thru.

Содержание 3656C W1

Страница 1: ...th 3743E EX mmWave Module VectorStar ME7838A4 A4X Modular Broadband VNA System with 3743A AX mmWave Modules VectorStar ME7838E4 E4X Modular Broadband VNA System with 3743E EX mmWave Modules VectorStar...

Страница 2: ......

Страница 3: ...Configuring the Hardware 2 1 2 4 W1 Verification Application Installation 2 2 2 5 Running the Verification Software 2 5 Procedure 2 5 Enter Port 1 BB mmWave Module Information 2 8 Enter Port 2 BB mmW...

Страница 4: ...E4 E4X 3 1 Introduction 3 1 3 2 Required Equipment 3 1 3 3 Configuring the Hardware 3 1 3 4 W1 Verification W1 Verification Application Installation 3 2 3 5 Running the Verification Software 3 5 Proc...

Страница 5: ...sts 3 32 Procedure 3 32 3 12 Mismatched Thru Tests 3 35 Procedure 3 35 3 13 Saving Verification Data 3 38 3 14 Low Band Calibration Port 3 Port 4 3 39 Procedure 3 39 3 15 High Band Calibration Port 3...

Страница 6: ...ab 4 10 4 7 Low Band Calibration 4 17 Procedure 4 17 4 8 High Band Calibration 4 20 Procedure 4 20 4 9 Application Interface Main Menu Tab 4 23 Auto Scale Display Button 4 24 Default Display Scale But...

Страница 7: ...on Causes of Connector Failure 6 1 Coaxial Connector Care 6 1 6 2 Visual Inspection 6 2 Outer Conductor and Thread Inspection 6 2 Center Conductor Inspection 6 3 6 3 Connector Cleaning 6 4 Required Cl...

Страница 8: ...Contents 6 PN 10410 00784 Rev A 3656C W1 Cal Ver Kit 2300 616 R PVS UG...

Страница 9: ...e used with other VNA systems including the ME7838D and ME7838D4 with appropriate adapters to the 1 mm connector interface The verification processes only apply to the systems enumerated above This ch...

Страница 10: ...2 02 mm 23W50 1 6 Female Offset Short 1 2 02 mm 23WF50 1 7 Male Offset Short 2 2 65 mm 23W50 2 8 Female Offset Short 2 2 65 mm 23WF50 2 9 Male Offset Short 3 3 180 mm 23W50 3 10 Female Offset Short 3...

Страница 11: ...ications Data Output The test data and results are output in the form of four files to a directory X installed directory on your computer s hard drive The default file names depending on the type of t...

Страница 12: ...lts and data are stored in four files on the computer hard disk drive These files are saved in ASCII format for easy viewing and printing at a later time 1 4 Related Documentation For more information...

Страница 13: ...test set and the VNA Necessary front panel interconnect cables between the test set and the VNA Necessary interconnect cables to the 3743x mmWave Modules 2 3 Configuring the Hardware This section desc...

Страница 14: ...n Windows Explorer browse to the USB Drive and double click on Startup htm 4 On the navigation page click the Install Anritsu 1 0 mm Verification Application Software link 5 The first of several insta...

Страница 15: ...ons Error 1321 Select Exit Installation and make sure you have Administrator privileges on the PC Controller Error 1931 Select OK to bypass the error message and continue the installation 8 Two instal...

Страница 16: ...ears Click Next and the Installation Complete dialog box appears 10 The installer adds all required files and makes the necessary registry updates on the PC Controller When done the navigation page Fi...

Страница 17: ...ollows 1 Choose your system configuration for Calibration Verification kit type and system model number 2 Measure Low Band a Calibrate the low frequency band b Measure verification devices for low fre...

Страница 18: ...nter a user name or other identification 5 Click OK to continue The program searches for a compatible VNA on the GPIB It will also determine if the system is a 2 Port or 4 Port system 6 When a VNA is...

Страница 19: ...yes then a Configure Application Settings dialog box appears 8 Select the calibration kit model and the system model number to be used Click Next to proceed The program will request the model number...

Страница 20: ...v A 3656C W1 Cal Ver Kit 2300 616 R PVS UG Enter Port 1 BB mmWave Module Information 9 Enter the Port 1 BB mmWave Module model number and click OK Figure 2 9 Anritsu W1 Verification Dialog Port 1 BB m...

Страница 21: ...Rev A 2 9 Enter Port 2 BB mmWave Module Information 10 Enter the Port 2 BB mmWave Module model number and click OK 11 Enter the Port 2 BB mmWave Module model number and click OK Figure 2 11 Anritsu W...

Страница 22: ...on 12 Enter the Test Set model number and click OK 13 Enter the Test Set serial number found on the rear panel and click OK 14 The program execution continues to the following Section 2 6 Application...

Страница 23: ...this section The Main Menu tab display is described in Section 2 9 Application Interface Main Menu Tab on page 2 25 The Serial Number Info tab is described in Section 2 10 Application Interface Serial...

Страница 24: ...ars 5 The VNA displays an acknowledgment dialog box when the file load is complete indicating that 10 files were loaded After the VNA has loaded the 10 Calibration coefficient files select OK on the P...

Страница 25: ...ect either the low range 40 MHz to 67 GHz or the high range 67 GHz to 110 125 GHz For purposes of this document the low band is done first The range that is not selected is calibrated during the Resta...

Страница 26: ...616 R PVS UG 7 In the Verification Kit pane select the W Coax type enter the Verification Kit Serial Number usually the same as the Calibration Kit When the Serial Number is complete the Locate USB D...

Страница 27: ...56C W1 Cal Ver Kit 2300 616 R PVS UG PN 10410 00784 Rev A 2 15 8 In the Locate USB Drive Kit Data on PC panel click the Go To USB Drive Locator button Figure 2 20 ME7838A Application Interface Setup M...

Страница 28: ...Kit button This utility allows you to manually select the path where the verification kit data is located The data path will be on the USB drive shipped with the verification kit On most PC s this wil...

Страница 29: ...6C W1 Cal Ver Kit 2300 616 R PVS UG PN 10410 00784 Rev A 2 17 10 Click Verify Files on USB Drive The application checks the files on the USB drive to verify that all files are present and that each ha...

Страница 30: ...nd you are ready to start click the Setup Calibration of VNA button 12 The program execution continues in either Section 2 7 Low Band Calibration on page 2 19 or Section 2 8 High Band Calibration on p...

Страница 31: ...ort Short Short Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 1 so that a Female connector is available to the operator b Inst...

Страница 32: ...NA measures the broadband load device data then the VNA Calibration Step 3 of 5 dialog box appears a Connect the Male Cal Kit Open to VNA Port 1 b Connect the Female Short 1 2 02 mm to VNA Port 2 c Wh...

Страница 33: ...A Calibration Step 5 of 5 dialog box appears a Remove the Short and the Open from Port 1 and Port 2 respectively b Slide the two T R modules towards each other and connect the VNA Port 1 Female to VNA...

Страница 34: ...appears a Install a F F adapter on VNA Port 1 so that a Female connector is available to the operator b Install a F M adapter on VNA Port 2 so that a Male connector is available to the operator c When...

Страница 35: ...e 2 65 mm Short 2 to VNA Port 1 b Connect the Female 3 18 mm Short 3 to VNA Port 2 c When ready to proceed click OK 5 The VNA Calibration Step 4 of 5 dialog box appears a Connect the Male 3 18 mm Shor...

Страница 36: ...p 5 of 5 dialog box appears a Remove the offset Shorts from Port 1 and Port 2 b Slide the two T R modules towards each other and connect the VNA Port 1 Female to VNA Port 2 Male directly c When ready...

Страница 37: ...Tab on page 2 11 The Main Menu tab display is described in the section immediately following and is shown in Figure 2 36 ME7838A Application Interface Main Menu Tab on page 2 25 The Serial Number tab...

Страница 38: ...program auto scales all four channels before continuing The VNA display during the calibration and verification tests is set to one Channel and four traces Each trace displays the following parameters...

Страница 39: ...nstrument the frequency range or the selected verification kit Selecting Yes causes the program to return to the program initialization stage Usually this button is used to test for second frequency b...

Страница 40: ...e 2 36 The number of saved files varies depending on the user settings If both the matched and mismatched thru tests are run two files are written to the PC Controller hard drive If only one of the te...

Страница 41: ...plication Interface Serial Number Info Tab The Serial Number Info tab displays a summary of the model number and serial number for the following devices All screen information is included in the repor...

Страница 42: ...x appears 2 Ensure that the correct device is connected The Matched Thru verification standard is the device with one scribe line The Mismatched Thru verification standard is the device with two scrib...

Страница 43: ...cted in the Main Menu tab above The VNA main display should be similar to the one below This is a typical display immediately after measuring the Matched Thru in the Low Band Please note this is for r...

Страница 44: ...nly the Matched Thru button was selected the program returns to the Application Interface Main Menu Tab on page 2 25 The dialog box is shown in Figure 2 36 ME7838A Application Interface Main Menu Tab...

Страница 45: ...page 2 25 The dialog box is shown in Figure 2 36 ME7838A Application Interface Main Menu Tab on page 2 25 If only the Mismatched Thru button was selected above program execution starts here If the Run...

Страница 46: ...the device label facing up 5 Connect the Mismatched Thru Airline female connector to VNA Port 2 6 When ready to continue click OK The Mismatched Thru Test Sweep 1 dialog box is displayed first followe...

Страница 47: ...0 616 R PVS UG PN 10410 00784 Rev A 2 35 typical display immediately after measuring the Mismatched Thru in the Low Band Please note this is for reference only and that the actual data on the VNA disp...

Страница 48: ...he main menu as described previously in Section 2 9 Application Interface Main Menu Tab on page 2 25 The default directory path and file names are C mmdcsvc Data Matched Thru Low dat C mmdcsvc Data Mi...

Страница 49: ...been saved to the C mmdscvc Data 2 To perform the verification tests for the second frequency range on the System Verification Software Main Menu Main Menu Tab select the Restart button The main menu...

Страница 50: ...k to see that the Windows GPIB is present on the boot drive that it is properly configured and that it passes the National Instruments hardware and software tests 3 This version of the verification so...

Страница 51: ...ng and in case you require factory assistance 4 When installing calibration devices and when measuring verification devices pay particular attention to proper connector alignment and torque Torque the...

Страница 52: ...2 15 Troubleshooting Using the PVS Software with the ME7838A AX E EX 2 40 PN 10410 00784 Rev A 3656C W1 Cal Ver Kit 2300 616 R PVS UG...

Страница 53: ...nnect cables between the test set and the VNA Necessary front panel interconnect cables between the test set and the VNA Necessary interconnect cables to the 3743x mmWave Modules 3 3 Configuring the H...

Страница 54: ...ot 3 Open Windows Explorer browse to the USB Drive and double click on Startup htm 4 On the navigation page click the Install Anritsu 1 0 mm Verification Application Software link 5 The first of sever...

Страница 55: ...uctions Error 1321 Select Exit Installation and make sure you have Administrator privileges on the PC Controller Error 1931 Select OK to bypass the error message and continue the installation 8 Two in...

Страница 56: ...ears Click Next and the Installation Complete dialog box appears 10 The installer adds all required files and makes the necessary registry updates on the PC Controller When done the navigation page Fi...

Страница 57: ...ication kit type and system model number 2 Select a two port pair to measure Ports 1 2 or 3 4 3 Measure Low Band on Ports 1 2 a Calibrate the low frequency band b Measure verification devices for low...

Страница 58: ...2 Locate the Anritsu W1 Verification icon shown below on the desktop and double click it 3 The program displays an About dialog box with version information Click OK to continue 4 The Operator Name di...

Страница 59: ...o continue or click No to search for another VNA 8 If Yes is clicked above a Port Selection dialog box appears 9 Select the calibration kit model and system model number to be used and click Next Figu...

Страница 60: ...d on only two ports at a time At the prompt below select to test either Port 1 and Port 2 or test Port 3 and Port 4 the click Next The program will request the model numbers and serial numbers of the...

Страница 61: ...4 Rev A 3 9 12 Enter the Port 1 BB mmWave Module serial number and click OK Enter Port 2 BB mmWave Module Information 13 Enter the Port 2 BB mmWave Module model number and click OK Figure 3 11 Anritsu...

Страница 62: ...ts 3 and 4 Then Enter Port 3 BB mmWave Module Information 15 Enter the Port 3 BB mmWave Module model number and click OK 16 Enter the Port 3 BB mmWave Module serial number and click OK Figure 3 13 Anr...

Страница 63: ...Rev A 3 11 Enter Port 4 BB mmWave Module Information 17 Enter the Port 4 BB mmWave Module model number and click OK 18 Enter the Port 4 BB mmWave Module serial number and click OK Figure 3 16 Anritsu...

Страница 64: ...irst cycle through the application will ask for the test set information for both Ports 1 and 2 and Ports 3 and 4 Enter Port 1 and Port 2 Test Set Information 20 Enter the Test Set model number and cl...

Страница 65: ...22 Enter the Test Set model number and click OK 23 Enter the Test Set serial number found on the rear panel and click OK The program execution continues to the following Section 3 6 Application Inter...

Страница 66: ...ain Menu tab display is described in Section 3 9 Application Interface Main Menu Tab on page 3 27 The Serial Number Info tab is described in Section 3 10 Application Interface Serial Number Tabs on pa...

Страница 67: ...into the VNA then click OK 6 The VNA displays an acknowledgment dialog box when the file load is complete indicating that 10 files were loaded After the VNA has loaded the 10 Calibration coefficient f...

Страница 68: ...67 GHz to 110 125 GHz For purposes of this document the low band is done first The range that is not selected is calibrated during the Restart Procedure described in Section 3 21 Restart for Alternat...

Страница 69: ...8 Select the W1 Coax type and then the Verification Kit Serial Number usually the same as the Calibration Kit When the Serial Number is complete the Locate USB Drive Kit Data on PC area appears 9 In...

Страница 70: ...rification Data Kit button This utility allows you to manually select the path where the verification kit data is located The data path will be on the USB drive shipped with the verification kit On mo...

Страница 71: ...656C W1 Cal Ver Kit 2300 616 R PVS UG PN 10410 00784 Rev A 3 19 11 Click Verify Files on USB Drive The application checks the files on the USB drive to verify that all files are present and that each...

Страница 72: ...a low band calibration and a high band calibration to be performed on only two ports at a time The program execution for port 1 and port 2 continues in either Section 3 7 Low Band Calibration Port 1 P...

Страница 73: ...a full 12 term SSST Short Short Short Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 1 so that a Female connector is available...

Страница 74: ...A Port 2 c When ready to proceed click OK 4 The VNA measures the broadband load device data then the VNA Calibration Step 3 of 5 dialog box appears a Connect the Male Cal Kit Open to VNA Port 1 b Conn...

Страница 75: ...VNA Calibration Step 5 of 5 dialog box appears a Remove the Short and the Open from Port 1 and Port 2 respectively b Slide the two T R modules towards each other and connect the VNA Port 1 Female to V...

Страница 76: ...rm SOLT Short Open Load Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 1 so that a Female connector is available to the operato...

Страница 77: ...Connect the Male 2 02 mm Short 1 to VNA Port 1 b Connect the Female 2 65 mm Short 2 to VNA Port 2 c When ready to proceed click OK 4 The VNA Calibration Step 3 of 5 dialog box appears a Connect the M...

Страница 78: ...ort 1 to VNA Port 2 c When ready to proceed click OK 6 The VNA Calibration Step 5 of 5 dialog box appears a Remove the offset Shorts from Port 1 and Port 2 b Slide the two T R modules towards each oth...

Страница 79: ...Setup Menu Tab on page 3 14 The Main Menu tab display is described in the section immediately following and is shown in Figure 3 43 The Serial Number tab is described in Section 3 10 Application Inter...

Страница 80: ...he program auto scales all four channels before continuing The VNA display during the calibration and verification tests is set to one Channel and four traces Each trace displays the following paramet...

Страница 81: ...o change the tested instrument the frequency range or the selected verification kit Selecting Yes causes the program to return to the program initialization stage Usually this button is used to test f...

Страница 82: ...38 The number of saved files varies depending on the user settings If both the matched and mismatched thru tests are run two files are written to the PC Controller hard drive If only one of the tests...

Страница 83: ...tion Interface Serial Number Tabs The Serial Number Info tab displays a summary of the model number and serial number for the following devices All screen information is included in the report headers...

Страница 84: ...b on page 3 27 Procedure 1 The Enter Serial Number Match Thru dialog box appears 2 Ensure that the correct device is connected The Matched Thru verification standard is the device with one scribe line...

Страница 85: ...Thru male connector to VNA Port 1 with the label facing up 6 Connect the Matched Thru female connector to VNA Port 2 7 When ready to proceed click OK The matched thru test starts and the Matched Thru...

Страница 86: ...l data on the VNA display may not be identical 9 If only the Matched Thru button was selected the program returns to the Application Interface Main Menu Tab on page 3 27 The dialog box is shown in Fig...

Страница 87: ...page 3 27 The dialog box is shown in Figure 3 43 ME7838A4 Application Interface Main Menu Tab on page 3 27 If only the Mismatched Thru button was selected above program execution starts here If the Ru...

Страница 88: ...h the device label facing up 5 Connect the Mismatched Thru Airline female connector to VNA Port 2 6 When ready to continue click OK The Mismatched Thru Test Sweep 1 dialog box is displayed first follo...

Страница 89: ...2300 616 R PVS UG PN 10410 00784 Rev A 3 37 typical display immediately after measuring the Mismatched Thru in the Low Band Please note this is for reference only and that the actual data on the VNA d...

Страница 90: ...d previously in Section 3 9 Application Interface Main Menu Tab on page 3 27 The default directory path and file names are C mmdcsvc Data Matched Thru Low dat C mmdcsvc Data Mismatched Thru Low dat C...

Страница 91: ...th a full 12 term SSST Short Short Short Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 3 so that a Female connector is availab...

Страница 92: ...ale Cal Kit Broadband Load Device to VNA Port 3 b Connect the Female Broadband Load Device to VNA Port 4 c When ready to proceed click OK 4 The VNA measures the broadband load device data then the VNA...

Страница 93: ...Port 4 c When ready to proceed click OK 6 The VNA measures the short open device data then the VNA Calibration Step 5 of 5 dialog box appears a Remove the Short and the Open from Port 3 and Port 4 res...

Страница 94: ...term SOLT Short Open Load Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 3 so that a Female connector is available to the opera...

Страница 95: ...Connect the Male 2 02 mm Short 1 to VNA Port 3 b Connect the Female 2 65 mm Short 2 to VNA Port 4 c When ready to proceed click OK 4 The VNA Calibration Step 3 of 5 dialog box appears a Connect the M...

Страница 96: ...ort 1 to VNA Port 4 c When ready to proceed click OK 6 The VNA Calibration Step 5 of 5 dialog box appears a Remove the offset Shorts from Port 3 and Port 4 b Slide the two T R modules towards each oth...

Страница 97: ...tup Menu Tab on page 3 14 The Main Menu tab display is described in the section immediately following and is shown in Figure 3 43 The Serial Number tab is described in Section 3 17 Application Interfa...

Страница 98: ...s is MAIN Scale SCALE Auto Scale Active Channel MAIN Scale SCALE Auto Scale All Channels When the program detects the VNA has finished a sweep the program auto scales all four channels before continui...

Страница 99: ...cal Default Display Scale Button This command sets all four channels to a scale of 5 dB div for the Magnitude displays and to 40 degrees div for the Phase displays and the reference value to 0 This al...

Страница 100: ...in Figure 3 70 Selecting Yes exits the program selecting No returns you to the previous screen About Displays version information copyright other legal notices and company contact information Run All...

Страница 101: ...the tests was run only one file is written to the PC Controller s hard drive at C mmcsvc Data Matched Thru This command button only runs the Matched Thru tests If selected program execution continues...

Страница 102: ...tion Interface Serial Number Tabs The Serial Number Info tab displays a summary of the model number and serial number for the following devices All screen information is included in the report headers...

Страница 103: ...b on page 3 45 Procedure 1 The Enter Serial Number Match Thru dialog box appears 2 Ensure that the correct device is connected The Matched Thru verification standard is the device with one scribe line...

Страница 104: ...r to VNA Port 4 7 When ready to proceed click OK The matched thru test starts and the Matched Thru Test Sweep 1 dialog box appears After it completes the Matched Thru Test Sweep 2 dialog box appears 8...

Страница 105: ...ogram returns to the Application Interface Main Menu Tab on page 3 45 The dialog box is shown in Figure 3 68 ME7838A4 Application Interface Main Menu Tab on page 3 45 10 If the Run All Tests button wa...

Страница 106: ...page 3 45 The dialog box is shown in Figure 3 68 ME7838A4 Application Interface Main Menu Tab on page 3 45 If only the Mismatched Thru button was selected above program execution starts here If the Ru...

Страница 107: ...the device label facing up 5 Connect the Mismatched Thru Airline female connector to VNA Port 4 6 When ready to continue click OK The Mismatched Thru Test Sweep 1 dialog box is displayed first follow...

Страница 108: ...he Mismatched Thru in the Low Band Please note this is for reference only and that the actual data on the VNA display may not be identical Figure 3 79 MS4640B Series VNA Typical Mismatched Thru Data i...

Страница 109: ...ction 3 16 Application Interface Main Menu Tab on page 3 45 The default directory path and file names are C mmdcsvc Data Matched Thru LowPort3and4 dat C mmdcsvc Data Mismatched Thru LowPort3and4 dat C...

Страница 110: ...l Restart button is used to select the alternative port pair in a 4 port system Procedure 1 Ensure that the verification results of the previous test have been saved to the C mmdscvc Data 2 To perform...

Страница 111: ...ernate Frequency Band or 3656C W1 Cal Ver Kit 2300 616 R PVS UG PN 10410 00784 Rev A 3 59 6 Dialog boxes will give the option to confirm the module and test set components or to correct the informatio...

Страница 112: ...ond set of calibration data and verification results to C mmdcsvc Data 9 This series of automated tests has verified that the VectorStar ME7838A4 A4X E4 E4X Broadband S parameter measurement accuracy...

Страница 113: ...sk for the Vector Network Analyzer support engineer for further assistance Difficulty Meeting System Specifications If the verification software appears to run properly but the results are not within...

Страница 114: ...3 22 Troubleshooting Using the PVS Software with the ME7838A4 A4X E4 E4X 3 62 PN 10410 00784 Rev A 3656C W1 Cal Ver Kit 2300 616 R PVS UG...

Страница 115: ...gnal Generators 2 to 20 GHz for LO and RF Sources 4 3 Configuring the Hardware This section describes how the various system elements are interconnected and the preliminary steps required for operatio...

Страница 116: ...xplorer browse to the USB Drive and double click on Startup htm 4 On the navigation page click the Install Anritsu 1 0 mm Verification Application Software link 5 The first of several installation dia...

Страница 117: ...Error 1321 Select Exit Installation and make sure you have Administrator privileges on the PC Controller Error 1931 Select OK to bypass the error message and continue the installation 8 Two installati...

Страница 118: ...Click Next and the Installation Complete dialog box appears 10 The installer adds all required files and makes the necessary registry updates on the PC Controller When done the navigation page Figure...

Страница 119: ...generator s rear panel The application process flow for a full verification of an ME7828A is as follows 1 Choose your system configuration for Calibration Verification kit type and system model number...

Страница 120: ...on Click OK to continue 4 The Operator Name dialog box appears Enter a user name or other identification 5 Click OK to continue The program searches for a compatible VNA on the GPIB 6 When a VNA is fo...

Страница 121: ...tion Settings dialog box appears 8 Select the calibration kit and system model number to be used When Next is clicked above the program searches for the Signal Generator LO Source and when found the L...

Страница 122: ...he rear panel of the upper Signal Generator and click OK 11 Enter the Source 2 RF model number lower Signal Generator usually an MG37022A and click OK 12 Enter the Source 2 RF serial number found on t...

Страница 123: ...nter the test set serial number found on the rear panel and click OK 14 Confirm that the test serial number is correct by clicking Yes 15 The program execution continues to the following Section 4 6 A...

Страница 124: ...in Menu tab display is described in Section 4 9 Application Interface Main Menu Tab on page 4 23 The Serial Number tab is described in Section Serial Number Tab Functions on page 4 26 The following pr...

Страница 125: ...he VNA then click OK 6 The VNA displays an acknowledgment dialog box when the file load is complete indicating that 10 files were loaded After the VNA has loaded the 10 Calibration coefficient files s...

Страница 126: ...the low range 40 MHz to 67 GHz or the high range 67 GHz to 110 GHz For purposes of this document the low band is done first The range that is not selected is calibrated during the Restart Procedure de...

Страница 127: ...A 4 13 8 Select the W Coax type and then the Verification Kit Serial Number usually the same as the Calibration Kit When the Serial Number is complete the Locate USB Drive Kit Data on PC area appears...

Страница 128: ...PN 10410 00784 Rev A 3656C W1 Cal Ver Kit 2300 616 R PVS UG 9 In the Locate USB Drive Kit Data on PC area click the Go To USB Drive Locator button Figure 4 20 ME7828A Application Interface Setup Menu...

Страница 129: ...it button This utility allows you to manually select the path where the verification kit data is located The data path will be on the USB drive shipped with the verification kit On most PC s this will...

Страница 130: ...that each has the proper number of data points 12 When all information is entered and you are ready to start click the Setup Calibration of VNA button 13 The program execution continues in either Sect...

Страница 131: ...Short Short Thru calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 1 so that a Female connector is available to the operator b Install...

Страница 132: ...easures the broadband load device data then the VNA Calibration Step 3 of 5 dialog box appears a Connect the Male Cal Kit Open to VNA Port 1 b Connect the Female Short 1 2 02 mm to VNA Port 2 c When r...

Страница 133: ...libration Step 5 of 5 dialog box appears a Remove the Short and the Open from Port 1 and Port 2 respectively b Slide the two T R modules towards each other and connect the VNA Port 1 Female to VNA Por...

Страница 134: ...u calibration Procedure 1 The W1 Verification Install Adapter dialog box appears a Install a F F adapter on VNA Port 1 so that a Female connector is available to the operator b Install a F M adapter o...

Страница 135: ...K 4 The VNA Calibration Step 3 of 5 dialog box appears a Connect the Male 2 65 mm Short 2 to VNA Port 1 b Connect the Female 3 18 mm Short 3 to VNA Port 2 c When ready to proceed click OK 5 The VNA Ca...

Страница 136: ...of 5 dialog box appears a Remove the offset Shorts from Port 1 and Port 2 b Slide the two T R modules towards each other and connect the VNA Port 1 Female to VNA Port 2 Male directly c When ready to p...

Страница 137: ...Menu Tab on page 4 10 The Main Menu tab display is described in the section immediately following and is shown in Figure 4 35 ME7828A Application Interface Main Menu Tab on page 4 23 The Serial Number...

Страница 138: ...ram auto scales all four channels before continuing The VNA display during the calibration and verification tests is set to one Channel and four traces Each trace displays the following parameters Upp...

Страница 139: ...ument the frequency range or the selected verification kit Selecting Yes causes the program to return to the program initialization stage Usually this button is used to test for second frequency band...

Страница 140: ...e PC Controller hard drive If only one of the tests was run only one file is written to the PC Controller s hard drive at C mmcsvc Data Matched Thru This command button only runs the Matched Thru test...

Страница 141: ...e with the ME7828A 4 9 Application Interface Main Menu Tab 3656C W1 Cal Ver Kit 2300 616 R PVS UG PN 10410 00784 Rev A 4 27 Signal Generator Sources Figure 4 38 ME7828A Application Interface Serial Nu...

Страница 142: ...ears 2 Ensure that the correct device is connected The Matched Thru verification standard is the device with one scribe line The Mismatched Thru verification standard is the device with two scribe lin...

Страница 143: ...in the Main Menu tab above The VNA main display should be similar to the one below This is a typical display immediately after measuring the Matched Thru in the Low Band Please note this is for refer...

Страница 144: ...the Matched Thru button was selected the program returns to the Application Interface Main Menu Tab on page 4 23 The dialog box is shown in Figure 4 35 ME7828A Application Interface Main Menu Tab on p...

Страница 145: ...23 The dialog box is shown in Figure 4 35 ME7828A Application Interface Main Menu Tab on page 4 23 If only the Mismatched Thru button was selected above program execution starts here If the Run All T...

Страница 146: ...e dialog box appears 4 Connect the Mismatched Thru Airline male connector to VNA Port 1 with the device label facing up 5 Connect the Mismatched Thru Airline female connector to VNA Port 2 6 When read...

Страница 147: ...nu described Section 4 9 Application Interface Main Menu Tab on page 4 23 and the display should look similar to Figure 4 46 This is a typical display immediately after measuring the Mismatched Thru i...

Страница 148: ...ain menu as described previously in Section 4 9 Application Interface Main Menu Tab on page 4 23 The default directory path and file names are C mmdcsvc Data Matched Thru Low dat C mmdcsvc Data Mismat...

Страница 149: ...e main menu is described in Section 4 9 Application Interface Main Menu Tab on page 4 23 3 The Choose The Restart Method dialog box appears as shown below in Figure 4 49 4 If you selected Full Restart...

Страница 150: ...y contact your Anritsu customer service center and ask for the Vector Network Analyzer support engineer for further assistance Difficulty Meeting System Specifications If the verification software app...

Страница 151: ...econds Terminate Read on EOS NO or unchecked Set EOI with EOS on Writes YES or checked Type of compare on EOS 8 bit EOS Byte 0Ah or decimal 10 Send EOI at End of Write YES or checked System Controller...

Страница 152: ...rameter Setting Primary GPIB Address 6 Secondary GPIB Address NONE Time out Setting 10 seconds Serial Poll Time out 1 second Terminate Read on EOS NO or unchecked Set EOI with EOS on Writes YES or che...

Страница 153: ...accuracy due to lack of repeatability Worn or damaged threads can damage mating connectors Over torquing connectors will introduce excessive stress to connector interface Under torquing will increase...

Страница 154: ...ors with bent broken or damaged contacts Magnification Use a device with a magnification range of 2X to 10X depending on the connector type Outer Conductor and Thread Inspection When connector threads...

Страница 155: ...Normal use Slight gap 3 Bent pin section Discard connector 4 Fingers spread or collapsed Discard connector 5 Broken pin or pin section Discard connector Figure 6 2 Example Connector Damage States Ind...

Страница 156: ...wabs can put lateral pressure on the center pin and damage it Inspect after cleaning to be sure that no cotton strands get caught in the connector Caution Teflon Tuning Washers The center conductor on...

Страница 157: ...n swab around the center pin Avoid applying pressure to the center pin When the connector is clean you should be able to hand tighten the connector to within approximately one half turn of the specifi...

Страница 158: ...as shown in Figure 6 9 on page 6 7 The male connector center pin must slip concentrically into the contact fingers of the female connector Never apply excessive force when making a connection 2 Push...

Страница 159: ...mage 3 Damaged pin Figure 6 9 Avoid Pin Damage Due to Misalignment Caution Do not turn the connector body Major damage to the center conductor and the outer conductor can occur if the connector body i...

Страница 160: ...size and rating for the connector and an open end wrench The open end wrench is used to prevent the body of the connector from turning Refer to Table 6 1 for torque specifications 2 Place the two wre...

Страница 161: ...ctor Type Size Wrench Size Torque Setting Recommended Tools 3 5 mma b a 3 5 mm 2 92 mm and K connectors are electrically compatible with SMA and have the same connector nut size 8 mm and torque requir...

Страница 162: ...50 67 110 Mating Type M F M F M F M F Sexless M F M F M F M F M F M F Cross mating Compatibility 7 16 DIN only Type N 20 GHz Type N 0 8 mm only 7 mm only 3 5 mm K K SMA 3 5 mm SMA V 2 4 mm W 1 1 mm on...

Страница 163: ......

Страница 164: ...Anritsu utilizes recycled paper and environmentally conscious inks and toner Anritsu Company 490 Jarvis Drive Morgan Hill CA 95037 2809 USA http www anritsu com...

Отзывы: