Using the PVS Software with the ME7838A/AX/E/EX
2-5 Running the Verification Software
3656C W1 Cal/Ver Kit & 2300-616-R PVS UG
PN: 10410-00784 Rev. A
2-5
2-5
Running the Verification Software
After the VNA and PC have been configured as described above, you are ready to run the program and perform
the calibration/verification. Make sure you have the following information available:
•
Serial number of the 3656B/C W1 Calibration/Verification Kit
•
Serial number of the 3739A/B/C Broadband/Millimeter Wave Test Set
•
Model number and serial number of the Port 1 BB/mm Module
•
Model number and serial number of the Port 2 BB/mm Module
The application process flow for a full verification of an ME7838A/AX/E/EX is as follows:
1.
Choose your system configuration for Calibration/Verification kit type and system model number.
2.
Measure Low Band
a.
Calibrate the low frequency band.
b.
Measure verification devices for low frequency band.
c.
Save verification data.
d.
Restart application for high frequency band.
3.
Measure High Band
a.
Calibrate the high frequency band.
b.
Measure verification devices for the high frequency band.
c.
Save verification data.
d.
Exit program.
Procedure
1.
With the equipment and software configured as described in
“Configuring the Hardware” on page 2-1
,
turn on the computer and allow it to start Windows.
2.
Locate the
Anritsu W1 Verification
icon (shown below) on the desktop and double-click it.
3.
The program displays an
About
dialog box with version information. Click
OK
to continue.
Note
Low-band (step 2) and high-band (step 3) calibration and verification can be performed in any order.
Figure 2-4.
About Dialog Box