16
Guidance and manufacturer´s declaration – electromagnetic
immunity –for EQUIPMENT and SYSTEM that are not
LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The JPD-FR100+ Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or the user of the
JPD-FR100+ Infrared Thermometer should assure that it is used in such an
environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
Portable and mobile RF
communications
quipment
should be used no closer to
any part of the J
Infrared Thermometer,
including cables, than the
recommended
separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
where p is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
b
Field strengths from fixed
RF transmitters, as
determined by an
Electromagnetic site
Содержание JPD-FR100+