Note: 

If your A2000/A2000I does not  have a SICK Module,                                                                    Service Procedure 

011 

          see Service Procedure 22.                                                                                                                    

(Page 1 of  3)  08/05/2014 

 

Level Sensor (SICK) Component Replacement

 

A2000/A2000I 

 
 
 
 
 
 
 
 
 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

1.  Turn the A2000 off. 
 
2.  The SICK module is blue and located in the top right of the Instrument. 
 
3.   Unscrew the SICK module from the bracket. 
 
4.  Use  the  screwdriver  to  pop  out  the  retaining  clip  that  holds  the  2  fiber  optic  cables  in  the 

socket.  Don’t lose the clip! 

 

 

5.  Remove the 2 fiber optic cables and insert them into the new SICK Module. 
 
6.  Pop the clip back in place to keep the cables from coming out.  The clip snaps in.  Make sure 

the fiber optic cables are fully inserted as the clip lines up with the groove on the cable ends. 

 
7.  There are 4 wires extending from the SICK module: #I 6 , #4, #3 and #3. 
 
 

 
 

SICK Module 

Clip 

Replacing the SICK Fiber Optic Module: 

ANKOM Technology  2052 O’Neil Rd   Macedon,  NY 14502  PH: (315) 986-8090   FAX: (315) 986-8091 