Note: 

If your A2000/A2000I does not  have a SICK Module,                                                                    Service Procedure 

011 

          see Service Procedure 22.                                                                                                                    

(Page 1 of  3)  08/05/2014 

 

Level Sensor (SICK) Component Replacement

 

A2000/A2000I 

 

ANKOM Technology  2052 O’Neil Rd   Macedon,  NY 14502  PH: (315) 986-8090   FAX: (315) 986-8091 

[email protected] 

Wire # I 6 

(White Wire) 

Green Relay 

(See 

step 14) 

Wire  #4, #3, #3

,  

9.  Remove the # I 6 wire from the gray wire terminal and replace it with the new wire 

from the new SICK module. 

 

10. Repeat  the  process  for  the  remaining  3  wires.  To  remove  a  wire  from  the  terminal 

block, insert  the provided screwdriver into the square hole directly next to  the wire 
you want to remove.  Don’t push too hard but it should snap in to release the wire.  
While  the  screwdriver  is  still  in  place,  insert  the  new  wire  with  the  same  number.  
Remove the screwdriver.  Lightly tug on each wire to test if it is securely in place. 

 
11. Attach  the  new  SICK  module  to  the  bracket.  If  you  have  an  older  unit,  the  SICK 

module attaches to the DIN rail next to the green relay. 

 
12.  Snap the bracket back in place to secure the module. 

 

13. Turn the A2000 on.  The module light should come on, indicating an EMPTY vessel. 

 

14. Open the vessel and gently clean off the glass tip with detergent/water using a cotton 

swab. Fill the vessel with water and test to see if the module light goes off when the 
glass tip is submerged in water.  You should hear the green relay “click” when the 
sensor detects water.  See Service Procedure # 6 Testing Level Sensor Section.  Do 
not retrain the level sensor.

 

 
 

 