Note: 

If your A2000/A2000I does not  have a SICK Module,                                                                    Service Procedure 

011 

          see Service Procedure 22.                                                                                                                    

(Page 1 of  3)  08/05/2014 

 

Level Sensor (SICK) Component Replacement

 

A2000/A2000I 

 

1.  Using a screw driver, remove the clip holding the fiber 

optic  cable  in  the  sensor.    Be  careful  not  to  lose  the 
clip. 

 
2.  Unplug the fiber optic cable. 
 
3.  Follow the cable up to the plastic fitting in the vessel. 
 
4.  Gently unscrew the plastic fitting.  (Allow the cable to 

turn freely.) 

5.  Replace the fitting/cable with the new fitting/

cable.  Be careful not to damage the glass tip 
protruding from the fitting. 

 

6. 

Tighten  the  fitting  firmly  by  hand  in  the 
vessel, then tighten the  fitting 1/4 turn  with a 
wrench  which  will  compress  the  O-ring  to 
make the seal.  

 
Reminder: 

The new glass sensor has to be used with the new 
fiber optic sending unit.  Do not mix and match the 
two components. 
 
7.  See  Test  the  Level  Sensor  Section  in  Service 

Procedure #6. 

 

 

ANKOM Technology  2052 O’Neil Rd   Macedon,  NY 14502  PH: (315) 986-8090   FAX: (315) 986-8091 

[email protected] 

Replacing the Glass Sensor Tip: 