DN:KD-5905-SMSY01 V1.0                                                                    Operation Guide 

Template  version:1 

8 / 16 

4. 

  CHANGE USER MEMORY SPACE

 

Two memories with 60 memory spaces each are available to save the measurement 

results of two different people  separately. In clock mode,  you can select the  desired 

user memory space by pressing the button 

“GROUP“. 

See picture 4 & 4-1.

 

       

 

Picture 4                    Picture 4-1

 

 

5.  CONNECTING THE CUFF TO THE MONITOR 

Insert the cuff tubing connector into the socket in the left side of the monitor. 

Make certain that the connector is completely inserted to avoid air leakage 

during blood pressure measurements.

 

  Avoid compression or restriction of the connection tubing during 

measurement

which may cause inflation error, or harmful injury due to 

continuous cuff pressure. 

 

6. 

  APPLYING THE CUFF

 

 

a.  Pulling the cuff end    through the medal loop (the cuff is packaged like 

this already), turn it outward (away from your body) and tighten it and 

close the Velcro fastener. 

b.  Place the cuff around a bare arm 1-2cm above the elbow joint.   

c.  While seated, place palm upside in front of you on a flat surface such 

as a desk or table. Position the air tube in the middle of your arm in 

line with your middle finger. 

d.  The  cuff  should  fit  comfortably,  yet  snugly  around  your  arm.  You 

should be able to insert one finger between your arm and the cuff.   

Note:

   

1. 

Please refer to the cuff circumference range in 

“SPECIFICATIONS” to make 

sure that the appropriate cuff is used.   

2. 

Measuring on same arm each time. 

3. 

Do not move your arm, body, or the monitor and do not move the rubber tube 

during measurement. 

4. 

Stay quiet, calm for 5 minutes before blood pressure measurement. 

5. 

Please keep the cuff clean. If the cuff becomes dirty, remove it from the monitor 

and clear it by hand in a mild detergent, then rinse it thoroughly in cold water. 

Never dry the cuff in clothes dryer or iron it. Clean the cuff after the usage of 

every 200 times is recommended.   

Содержание KD-5905

Страница 1: ...NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 6 3 VOICE SETTING 7 4 CHANGE USER MEMORY SPACE 8 5 CONNECTING THE CUFF TO THE MONITOR 8 6 APPLYING THE CUFF 8...

Страница 2: ...t The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peo...

Страница 3: ...EC 60601 1 2005 EN 60601 1 2006 AC 2010 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electric...

Страница 4: ...uring on same arm for each measurement 6 Please always relax at least 1 or 1 5 minutes between measurements to allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure...

Страница 5: ...other infective person to avoid cross infection 16 Medical AC adapter which output is DC 6 0V 600mA and complied with IEC 60601 1 EN 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 UL 60601 1 2 is s...

Страница 6: ...regulations at the end of their usage 2 CLOCK AND DATE ADJUSTMENT Radio Controlled Clock RCC Receiving a Initial receiving Once you install the battery the machine will enter the RCC Receiving Mode T...

Страница 7: ...number is blinking press the button MEM to increase the number Keep on pressing the button MEM the number will increase fast c You can exit the Clock Adjusting Mode by pressing button START when the...

Страница 8: ...turn it outward away from your body and tighten it and close the Velcro fastener b Place the cuff around a bare arm 1 2cm above the elbow joint c While seated place palm upside in front of you on a f...

Страница 9: ...segment is missing b Then the most recent result will be displayed with date and time See picture 8 1 Picture 8 Picture 8 1 Picture 8 2 c If the voice function is switched on the monitor will speak ou...

Страница 10: ...o result stored LCD will show See picture 9 2 Picture 9 Picture 9 1 Picture 9 2 c If the voice function is on the monitor will announce each result displaying on the screen d When displaying the store...

Страница 11: ...conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of blood pressure 12 TROUBLESHOOTING 1 PROBLEM POSSIBLE CAUSE SOLUT...

Страница 12: ...c interference Take out batteries for five minutes and then reinstall all batteries MAINTENANCE 1 Do not drop this monitor or subject it to strong impact 2 Avoid high temperature and solarization Do n...

Страница 13: ...applied part Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local Authority or retaile...

Страница 14: ...applicable Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic immunity The KD 5905 is intended for use in the electromagnetic environment specified be...

Страница 15: ...may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electrom...

Страница 16: ...quipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 3...

Отзывы: