DN:KD-5905-SMSY01 V1.0                                                                    Operation Guide 

Template  version:1 

4 / 16 

10.  Measurement range: 

Cuff pressure: 0-300mmHg 

Systolic:          60-260mmHg 

Diastolic:        40-199mmHg 

Pulse rate:      40-180 beats/minute 

11.  Accuracy:   

Pressure:        ±3mmHg 

Pulse rate:      ±5% 

12.  Environmental temperature for operation: 5

℃~

40

(41

℉~

104

13.  Environmental humidity for operation: 

90%RH 

14.  Environmental temperature for storage and transport: -20

℃~

55

(-4

℉~

131

15.  Environmental humidity for storage and transport: 

95%RH 

16.  Environmental pressure: 80KPa-105KPa 

17.  Battery life: Approx 300 times. 

18.  A  list  of  all  components  belonging  to  the  pressure  measuring  system,  including 

accessories: Pump,Valve, LCD, Cuff, Sensor 

 

Note: These specifications are subject to change without notice. 

 

NOTICE   
 

1.  Read all of the information in the operation guide and any other literature in the box before 

operating the unit. 

2.  Stay still, calm and rest for 5 minutes before blood pressure measurement. 

3.  The cuff should be placed at the same level as your heart. 

4.  During measurement, neither speak nor move your body and arm. 

5.  Measuring on same arm for each measurement. 

6.  Please always relax at least 1 or 1.5 minutes between measurements to allow the blood 

circulation  in  your  arm  to  recover.  Prolonged  over-inflation  (cuff  pressure  exceed  300 

mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause 

ecchymoma of your arm. 

7.  Consult your physician if you have any doubt about below cases: 

1) The application of the cuff over a wound or inflammation diseases; 

2) The application of the cuff on any limb where intravascular access or therapy, or an 

arterio-venous (A-V) shunt, is present; 

3) The application of the cuff on the arm on the side of a mastectomy; 

4) Simultaneously used with other monitoring medical equipments on the same limb; 

5) Need to check the blood circulation of the user. 

8. 

 

This  Electronic  Sphygmomanometers  is  designed  for  adults  and  should  never  be 

used  on  infants  or  young  children.  Consult  your  physician  or  other  health  care 

professionals before use on older children. 

9.  Do not use this unit in a moving vehicle, This may result in erroneous measurement. 

10.  Blood  pressure  measurements  determined  by  this  monitor  are  equivalent  to  those 

obtained by a trained observer using the cuff/stethoscope auscultation method, within the 

Содержание KD-5905

Страница 1: ...NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 6 3 VOICE SETTING 7 4 CHANGE USER MEMORY SPACE 8 5 CONNECTING THE CUFF TO THE MONITOR 8 6 APPLYING THE CUFF 8...

Страница 2: ...t The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peo...

Страница 3: ...EC 60601 1 2005 EN 60601 1 2006 AC 2010 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electric...

Страница 4: ...uring on same arm for each measurement 6 Please always relax at least 1 or 1 5 minutes between measurements to allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure...

Страница 5: ...other infective person to avoid cross infection 16 Medical AC adapter which output is DC 6 0V 600mA and complied with IEC 60601 1 EN 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 UL 60601 1 2 is s...

Страница 6: ...regulations at the end of their usage 2 CLOCK AND DATE ADJUSTMENT Radio Controlled Clock RCC Receiving a Initial receiving Once you install the battery the machine will enter the RCC Receiving Mode T...

Страница 7: ...number is blinking press the button MEM to increase the number Keep on pressing the button MEM the number will increase fast c You can exit the Clock Adjusting Mode by pressing button START when the...

Страница 8: ...turn it outward away from your body and tighten it and close the Velcro fastener b Place the cuff around a bare arm 1 2cm above the elbow joint c While seated place palm upside in front of you on a f...

Страница 9: ...segment is missing b Then the most recent result will be displayed with date and time See picture 8 1 Picture 8 Picture 8 1 Picture 8 2 c If the voice function is switched on the monitor will speak ou...

Страница 10: ...o result stored LCD will show See picture 9 2 Picture 9 Picture 9 1 Picture 9 2 c If the voice function is on the monitor will announce each result displaying on the screen d When displaying the store...

Страница 11: ...conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of blood pressure 12 TROUBLESHOOTING 1 PROBLEM POSSIBLE CAUSE SOLUT...

Страница 12: ...c interference Take out batteries for five minutes and then reinstall all batteries MAINTENANCE 1 Do not drop this monitor or subject it to strong impact 2 Avoid high temperature and solarization Do n...

Страница 13: ...applied part Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local Authority or retaile...

Страница 14: ...applicable Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic immunity The KD 5905 is intended for use in the electromagnetic environment specified be...

Страница 15: ...may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electrom...

Страница 16: ...quipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 3...

Отзывы: