ELECTROMAGNETIC COMPATIBILITY INFORMATION 

 

Table 1 

For all ME EQUIPMENT and ME SYSTEMS 

Guidance and manufacture’s declaration - electromagnetic emissions 

The [KD-5905] is intended for use in the electromagnetic environment specified below. 

The customer or the user of the [KD-5905] should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions 

CISPR 11 

Group 1 

The [KD-5905] uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to cause 

any interference in nearby electronic equipment. 

RF emissions 

CISPR 11 

Class B 

The [KD-5905] is suitable for use in all establishments other than 

domestic and those directly connected to the public low-voltage 

power supply network that supplies buildings used for domestic 

purposes. 

 

Harmonic emissions 

IEC 61000-3-2 

Not applicable 

Voltage fluctuations/ 

flicker emissions 

IEC 61000-3-3 

Not applicable 

 

Table 2 

For all ME EQUIPMENT and ME SYSTEMS 

Guidance and manufacturer’s declaration - electromagnetic immunity 

The [KD-5905] is intended for use in the electromagnetic environment specified below. The customer or the user of 

the [KD-5905] should assure that it is used in such an environment. 

IMMUNITY test 

IEC 60601test level 

Compliance level 

Electromagnetic environment - 

guidance 

Electrostatic 

discharge (ESD) 

IEC 61000-4-2 

± 6 kV contact 

± 8 kV air 

± 6 kV contact 

± 8 kV air 

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 

synthetic material, the relative humidity 

should be at least 30 %. 

Power frequency 

(50/60 Hz) 

magnetic field 

IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should 

be at levels characteristic of a typical 

location in a typical commercial or hospital 

environment. 

NOTE: U

T

 is the a.c. mains voltage prior to application of the test level. 

 

Table 3 

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Guidance and manufacturer’s declaration - electromagnetic immunity 

The [KD-5905] is intended for use in the electromagnetic environment specified below. The customer or the user 

of the [KD-5905] should assure that it is used in such an environment. 

Содержание KD-5905

Страница 1: ...NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 6 3 VOICE SETTING 7 4 CHANGE USER MEMORY SPACE 8 5 CONNECTING THE CUFF TO THE MONITOR 8 6 APPLYING THE CUFF 8...

Страница 2: ...t The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peo...

Страница 3: ...EC 60601 1 2005 EN 60601 1 2006 AC 2010 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electric...

Страница 4: ...uring on same arm for each measurement 6 Please always relax at least 1 or 1 5 minutes between measurements to allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure...

Страница 5: ...other infective person to avoid cross infection 16 Medical AC adapter which output is DC 6 0V 600mA and complied with IEC 60601 1 EN 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 UL 60601 1 2 is s...

Страница 6: ...regulations at the end of their usage 2 CLOCK AND DATE ADJUSTMENT Radio Controlled Clock RCC Receiving a Initial receiving Once you install the battery the machine will enter the RCC Receiving Mode T...

Страница 7: ...number is blinking press the button MEM to increase the number Keep on pressing the button MEM the number will increase fast c You can exit the Clock Adjusting Mode by pressing button START when the...

Страница 8: ...turn it outward away from your body and tighten it and close the Velcro fastener b Place the cuff around a bare arm 1 2cm above the elbow joint c While seated place palm upside in front of you on a f...

Страница 9: ...segment is missing b Then the most recent result will be displayed with date and time See picture 8 1 Picture 8 Picture 8 1 Picture 8 2 c If the voice function is switched on the monitor will speak ou...

Страница 10: ...o result stored LCD will show See picture 9 2 Picture 9 Picture 9 1 Picture 9 2 c If the voice function is on the monitor will announce each result displaying on the screen d When displaying the store...

Страница 11: ...conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of blood pressure 12 TROUBLESHOOTING 1 PROBLEM POSSIBLE CAUSE SOLUT...

Страница 12: ...c interference Take out batteries for five minutes and then reinstall all batteries MAINTENANCE 1 Do not drop this monitor or subject it to strong impact 2 Avoid high temperature and solarization Do n...

Страница 13: ...applied part Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local Authority or retaile...

Страница 14: ...applicable Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic immunity The KD 5905 is intended for use in the electromagnetic environment specified be...

Страница 15: ...may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electrom...

Страница 16: ...quipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 3...

Отзывы: