Instruction Manual
–
V6.19
GYROTORK DUAL version
- 14 -
3.
Type of Data
You can either choose to apply
the set point to the “Raw Data” or the “Peak value”.
Usually, the raw
data is used but you can setup a limit on the maximum or a calculation (mean, standard deviation,
break... see chapter “Graph calculation settings”).
4.
Top and Bottom limits
You can setup a top and a bottom limit for each channel by touching the button and entering the
limit value from the displayed keyboard.
Then you
can select “buzzer” if desired for each limit.
The set points will be displayed on the main measurement screen on the bar graph area and during
the test; the current reading will change of color when a limit is reached.
5.
Activate an Output
The limit can control a TTL output action. Select the desired output. The output action is chosen by
setting the desired output in the output control. Set the action and state as required.
Operations:
•
If the bottom limit is the only one set, the state of the output will change when the measured
value will exceed the limit value.
•
If the bottom and the top limit are set, the state of the output will change when the
measured value will be in-between the limit values