1.
Measurement principles
When the curve mode is activated, the torque gauge is under the
“TEST”
mode. Here is the process
of a TEST:
By default, the gauge graphs the curve of the active channel/Time for the Star version, and
channel1/channel2 for the Dual version. They are 3 values for the graph scale: 25%, 50%, and 100%.
At the end of the test, the graph is automatically drawn to efficiently use the space available.
There are 3 possibilities for the axis position: positive measure, negative measure, positive and
negative measure.
2.
States Indicators
On the left corner of the main screen, the state indicator is displayed:
Stand-by mode. Waiting for an event to start.
Acquisition mode. Acquiring data.
Test complete. To start a new test tares the torque gauge.
If you would like to stop the test before the end of the time, press the state indicator.
3.
Graph displaying area
To start measuring and drawing the graph, you must touch the main display area to tare. When the
GYROTORK detects a load on the sensor, it will start measuring torque & drawing the graph.
While applying a load to the gauge, the graph is drawn on the screen and the current and maximum
values are displayed according to the setup in the SCREENS settings.
The GYROTORK data capture rate is 5,000Hz, however, for better readability; it displays a maximum
of 1000 points per graph.
Tare torque
gauge
Waits to detect
torque on the
sensor
Start measuring,
recording torque
and drawing the
graph
Stop the test at
the end of the
set time.