Section 4 - Wiring
4.1 SIGNAL PROBE WIRING
Always route signal cables separate from AC power and AC control circuits. Signal cables must be shielded
with the bare shield drain wire grounded on one end and floating on the other end.
NOTE: If it becomes necessary to splice cable connections, be sure to properly solder or join wire
connections. Loose connections have the ability to generate false counts.
DIAGRAM 4.2 WIRING DIAGRAM
ANDERSON INSTRUMENT CO., INC.
PART NUMBER HT750000
MALE PIN (3)
WHITE WIRE
FEMALE SOCKET (1)
BLACK WIRE
WHITE
BLACK
SHIELD
PIN/SOCKET ORIENTATION
HT750000 PICK-UP PROBE SPECIFICATIONS
SIGNAL OUT
OPERATIONAL FREQUENCY RANGE
OUTPUT VOLTAGE
SINE WAVE
0-1000Hz
.18 - 2.6 V.P.P.
HT750000 LENGTH - APPROX. 12''
WHEN THREADED INTO HM SERIES FLOWMETER
HMP SERIES
TURBINE FLOWMETER
V
OUTPUT WIRING
Hz
SIGNAL OUTPUT
NO POLARITY
(
)
ANDERSON INSTRUMENT CO., INC.
RED + BLK COM WH SIG
PART NUMBER HTS1000
MALE PIN (3)
MALE PIN (2)
WHITE WIRE
RED WIRE
FEMALE SOCKET (1)
BLACK WIRE
RED (SUPPLY INPUT+8-24VDC)
WHITE (SIGNAL OUTPUT)
BLACK (COMMON)
SHIELD
PIN/SOCKET ORIENTATION
EXCITATION VOLTAGE
SIGNAL OUT
OPERATIONAL FREQUENCY RANGE
DUTY CYCLE
8-24 VDC, 0.8 Watt max
SQAURE WAVE
0-1000Hz
40 HIGH - 60 LOW
HTS1000 LENGTH - APPROX. 12''
WHEN THREADED INTO HM SERIES FLOWMETER
HMP SERIES
TURBINE FLOWMETER
V
SUPPLY VOLTAGE MINUS 0.7V
0.5V
INPUT/OUTPUT WIRING
T
~
~
40%
60%
40%