LifeSource Health Line (Toll-Free): 1-888-726-9966
E-2
WHAT DISPLAY SYMBOLS MEAN
?
Display
Symbol
Recommended
Action
Measurement is in progress,
remain as still as possible.
Take measurement again
and consult with your
physician.
Replace battery with new
one when the symbol blinks.
Alkaline battery recom-
mended.
Try the measurement again.
Remain very still during the
measurement.
Fasten the cuff correctly, and
try the measurement again.
Check for air leaks along the
tube and around the
air socket.
Refasten the cuff and retake
measurement.
Make sure tube is properly
connected to cuff and
monitor.
Try the measurement again.
Remain very still during the
measurement.
Press exhaust button to
release air.
Inflate cuff 30 to 40 mmHg
above your normal systolic
pressure.
Condition/Cause
The symbol shown while
measurement is in progress. It
blinks while detecting the
pulse.
An irregular heartbeat or
body movement may have
occurred. Refer to page E-9
for more information on
irregular heartbeats.
The battery power indicator
during measurement.
The battery is low when it
blinks.
Unstable blood pressure due
to excessive body movement.
The systolic and diastolic val-
ues are within 10 mmHg of
each other.
The pressure value did not
increase during inflation.
The cuff is not fastened
correctly.
There is an air leak in the cuff
or monitor.
The pulse is not detected
correctly.
Appears on display prior to
and during measurement.
Blinks if there is air in cuff at
power up.
Cuff needs to be inflated for
proper measurement.
Battery Full
PULSE DISPLAY ERROR
?
Battery Low