30
10. ID NUMBER AND GLP
The ID number is used to identify the balance when Good Laboratory Practice
(GLP) is used.
The ID number is stored in memory even if power to the balance is switched OFF.
The following GLP data is transmitted to a printer or a computer using the optional
RS-232C serial interface.
• The result of calibration (“Calibration report”)
• The result of calibration test (“Calibration test report”)
• The “Title block” and “End block” for GLP data
The GLP output format includes the balance manufacturer name, model number,
serial number, ID number and space for a signature.
When used with the AD-8121B, the date and time can be printed (GLP output
format
info
=
1
).
10-1. Setting the ID number
1. Press and hold the SAMPLE key to display
func
.
2. Press the SAMPLE key several times to display
id
.
3. Press the PRINT key. Enter the ID number using
the following keys.
SAMPLE key To select the digit blinking to be
changed.
RE-ZERO key To set the character of the digit
selected. See the table below for
the “display character set”.
PRINT key
To store the value and proceed to
the next step.
UNITS key
To cancel the value and proceed
to the next step.
4. When the above operation has been completed,
func
appears after
end
.
5. Press the UNITS key to return to the weighing
mode.
Display character set
0 1 2 3 4 5 6 7 8 9 - A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 - _ a b C d e f g H i j k l m n o p q r s t U v w x y z
“ ” : Space
Press
several times
func
id
000000
end
id
func
Set using the
relevant keys
Returns to
the weighing mode
Press and hold