Startup/operation
UV lamp for InnuPure C16
Edition 03/2013
15
Fig. 6
Inserting lamp into InnuPure C16, connecting and switch-on
6.2
Switching off the UV lamp for the InnuPure C16
1. Wait until the set operating time has elapsed and the UV lamp is switched off.
The green status LED at the front of the UV lamp turns off when the set oper-
ating time of the time relay has elapsed.
2. Disconnect the mains plug from the UV lamp and the mains socket.
3. Remove the UV lamp from the InnuPure C16.
CAUTION! Warming of the housing in the illumination area!
Depending on the operating time, the housing of the UV lamp may be heated up to 60° in the
illumination area.
6.3
Setting the operating time of the UV lamp with the time relay
Depending on the coding of the DIP switches (1, 2, 3), the operating time of the UV lamp can
be set with the dial. The time can be set from 0.05 s to 100 h. As an example, the table dis-
plays all possible codings for four setting values of the time relay. The factory setting of the
operating time is 30 min.