Commands for operation
AS 21hp, AS 10e
24
5
Commands for operation
5.1
Report-Commands
S
Show state Qxy
x:
Bit0
not used
y:
Bit0
error
Bit1
not used
Bit1
not used
Bit2
new switch on
Bit2
sampler stop
Bit3
busy
Bit3
not used
T
Identification rack Txy
x: 7
y: 1=21, 3=76, 4=54, 5=146, 6=20, 7=50 (dilution), 8=100
V
Version
Vx.yyy
F
Error code
Fxy
x:
Bit0 error rack actuator
y: Bit0
not used
Bit1 error rotate actuator
Bit1
not used
Bit2 error Z-actuator
Bit2
not used
Bit3 error identification rack
Bit3
not used
I
Initialize
Gx
Go to sample position
x = sample position
Tax
Putting down
x = steps (8 steps = 1 mm)
5.2
Adjustment commands
GJT
Go to Adjustment position
XRx
x = 1/10°
Rotate actuator counterclockwise
XLx
Rotate actuator clockwise
JT
Save position