Fault removal
InnuPure C16
Edition 06/2015
49
Maxdiff value Y1-axis exceeded
Maxdiff value Y2-axis exceeded
Maxdiff value Z-axis exceeded
Maxdiff value P1-axis exceeded
Maxdiff value M-axis exceeded
Maximum permissible position deviation exceeded in a
drive axis
Maxforce value Y1-axis exceeded
Maxforce value Y2-axis exceeded
Maxforce value Z-axis exceeded
Maxforce value P1-axis exceeded
Maxforce value M-axis exceeded
Permissible maximum current of a drive axis exceed-
ed
Difference temperature sensor val-
ues too high
Maximum permissible difference between the two
temperature sensor values exceeded
Plausibility error temperature sen-
sor left
Plausibility error temperature sen-
sor right
Temperature value of the left or right temperature
sensor not plausible
Error heating system
Error in the heating system
Timeout by communication
Timeout during the execution of an internal command
of the user interface
Device error!
Generally displayed after all error messages in an ad-
ditional message window.
7.2
General information and tips for preventing problems
Observe the following notes to prevent errors during the daily routine:
1.
After it is switched on the device automatically moves to its start position. Any fluids still
present in the tips after an abortion/power failure are not discharged in the process. In
this case, the fluid can be discharged using the tool [E
MPTYING
_T
IPS
] in the T
OOLS
menu
(see the section "Checking and executing individual device functions with tools" p. 36).
2.
Wait approx. 30 s after switching on the InnuPure C16 for the device initialization to be
completed, which is indicated by a beeping sound. Only then may the HID-Pro 320 be
operated or the sample tray be moved into the device. Otherwise, program faults may
result.
3.
Before starting an extraction protocol or a tool, make sure that no consumables from
previous processes have remained in the device. Remove any such consumables using
the tools [E
MPTYING
_T
IPS
], [W
ASTE
_T
IPS
] and [T
RAY
_O
UT
] in the T
OOLS
menu.
4.
If tips are present at the cones, no protocol may be started and the tools [A
UTOCALIBRA-
TION
] and [T
RANSPORT LOCK
] may not be executed. Otherwise the tips might be dam-
aged and the device be calibrated incorrectly, thereby causing more follow-on errors.
5.
The InnuPure C16 can be switched off at any time from the mains switch at the equip-
ment backplate, even while an extraction protocol is running. Switching off using the
touch sensor is only possible if this has been enabled.
6.
To protect against accidentally switching the device off, the touch sensor function is
disabled between calibration and the completion of the protocol when protocols/tools
are run during which fluids are moved. Since in the case of an aborted protocol or error
after calibration, the touch sensor function is not re-enabled automatically, this can be
performed by the user via [S
ETTINGS
]
[D
EVICE COMMANDS
]
[T
OUCH SENSOR ON
].