Operation
46
Edition 06/2015
InnuPure C16
Switching user login at start on/off
The automatic user login at the start of the InnuPure C16 is a program default and can be
switched on or off by users with administrator rights.
1.
Press the [M
ENU
]
button in the top left corner of the start screen.
2.
Enable the L
OGIN AT START
menu function if a login prompt should be shown when start-
ing the InnuPure C16, otherwise disable it the function.
If the L
OGIN AT START
function is disabled, the last logged-in user is automatically logged in at
the next program start.
Logging in a different user
While working in the program of the InnuPure C16 a different user can be logged in:
1.
Press the [M
ENU
]
button in the top left corner of the start screen.
2.
Select the menu item [N
EW LOGIN
].
3.
Enter the new user name and the corresponding password and confirm your entries
with [O
K
].
6.3.8
Enabling the interior light and the touch sensor
The interior of the InnuPure C16 is illuminated during the processing of an extraction proto-
col. However, the illumination can also be switched on and off separately.
During the processing of an extraction protocol the touch sensor is disabled to prevent the
device from being switched off accidentally. If a running protocol is aborted using the [S
TOP
]
button or as a result of an error, the touch sensor remains disabled and must be switched
back on if required:
1.
On the start screen press the [S
ETTINGS
] button.
2.
Select the menu item [D
EVICE COMMANDS
].
3.
Press the [L
IGHT ON
/
OFF
] button to switch the light on and back off.
4.
Press the [T
OUCH SENSOR ON
]
button to re-enable the touch sensor.
Figure 14 Switching the interior light/enabling the touch sensor