background image

Sub MkIIP Oxygen & Carbon Dioxide Monitor 

User Manual - US Navy Version 

 

 

Document Ref

: AS2-802-14 - 

January 2016 

Page 5 

 

 

Operation 

 

The external appearance of the SUB MkIIP is shown in Figure 1. 
 

 

Figure 1 - External View of Instrument 

 
Operation  of  the  SUB  MkIIP  is  simple,  the  only  controls  are  the  power  switch  and  a  single 
pushbutton  switch.    To  switch  the  unit  on,  turn  the  power  switch  to  either  the  'Battery'  or 
'External'  position.      Either  an  external  supply  or  batteries  must  be  installed  in  order  for  the 
unit to operate. 
 
If the unit does not turn on, perform the following checks: 
 
a) 

check that either the external power supply or batteries are healthy 

b) 

check that the fuses in the external supply line or the battery circuit are not blown 

 
On  powering  up  the  instrument,  the  sensor  readings  may  take  a  short  time  to  settle, 
particularly  the  CO

2

 

sensor,  which  can  take  about  40  seconds  to  ‘warm  up’.    During  this 

period, the CO

2

 

display shows “----“ to hide any erroneous measurements.  

 
Sensor readings are updated approximately every two seconds.    
 
The  choices  of  measurement  units  for  each  parameter  are  defined  at  the  time  of  order  and 
are factory set.  
 
Oxygen readings are displayed in %SEV (percent surface equivalent).   
 
CO

readings are displayed in %SEV (percent surface equivalent). 

 

SUB  Mk  II 
P

 

BACKLIGHT SWITCH

 

OXYGEN SENSOR

 

SUPPLY 
SWITCH

 

CARBON DIOXIDE 

 

SENSOR

 

BATTERY

 

CONDITION

 

ELAPSED 
TIME

 

MAINTAIN 

PRESSED 

TO

 

RESET ELAPSED TIME

 

BACKLIGHT

 

DD:H
H

 

EXT

 

POWER

 

OFF

 

ESTIMATED

 

HOURS

 

BATT

 

CARBON

 

DIOXIDE

 

TEMPERATUR
E

 

DEPTH

 

FSW

 

OXYGEN

 

%SEV

 

°F

 

%SEV

 

Содержание SA1 W75 SR 60 N 564

Страница 1: ...ck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net Sub MkIIP Oxygen Carbon Dioxide Monitor User Manual US Navy Version...

Страница 2: ......

Страница 3: ...al power 100 240VAC 8 6 Calibration 9 6 1 Readout check 9 6 2 Use of calibration gas 10 7 Data logging 11 7 1 Installation of data logging software on a PC 11 7 2 Data logging program overview 13 7 3...

Страница 4: ...Sub MkIIP Oxygen Carbon Dioxide Monitor User Manual US Navy Version Document Ref AS2 802 14 January 2016 Page 2...

Страница 5: ...supply cable Part Number SAEP AS2 605 c User manual d Test certificates e Analox hyperbaric tested batteries x4 Part Number 2557 4625 Optional f Data output cable Part Number SAEP AS2 602 and softwar...

Страница 6: ...ch the standard instrument has only one operator control A single pushbutton controls backlights for each of the liquid crystal displays LCD and can also be used to reset the elapsed time The unit is...

Страница 7: ...the fuses in the external supply line or the battery circuit are not blown On powering up the instrument the sensor readings may take a short time to settle particularly the CO2 sensor which can take...

Страница 8: ...nstrument is operating from an external power supply The backlight is turned on by momentarily pressing the pushbutton on the front of the instrument The backlight will turn off automatically after a...

Страница 9: ...nt damage open the lid of the instrument using the two catches beside the handle b Undo the two Velcro retaining straps around the batteries c Ease each battery from its clips use one hand to grip the...

Страница 10: ...ypical current consumption figures are shown below Input Voltage V DC Load Current without backlight mA Load Current with backlight mA 9 0 160 315 12 0 105 340 24 0 55 290 40 0 37 270 5 2 External pow...

Страница 11: ...the right hand column as a guideline to the acceptable tolerance Also ensure that the SUB MkIIP is sampling the atmosphere from the same compartment as the CAMS To convert the CAMS Gas reading from u...

Страница 12: ...ed with the instrument will be required and should be fitted as shown in Figure 2 The longer free pipe should be connected to the outlet of a test gas bottle flow regulator The flow adaptors in the ch...

Страница 13: ...ensure all data from a DISSUB event is captured A cable is supplied to connect the SUB MkIIP to the serial port on a PC The cable is fitted with a 9 way female D type connector An adaptor will also b...

Страница 14: ...ED THE NEXT TIME YOU START THE PROGRAM 11 Always ensure that the Baud rate is set to 9600 to which it will default 12 Click on the Communicate with Instrument button You will see a message Please wait...

Страница 15: ...iable that you wish to view and a graph will be automatically generated This process will suspend logging you will therefore need to click on the Restart logging button on the Instrument Status page T...

Страница 16: ...be apparent Full Memory This option enables you to specify what the instrument should do when its data logging memory is full You may choose to Ignore newer data if for example you are conducting a s...

Страница 17: ...spreadsheet Therefore select an Excel compatible csv file CSV stands for a Comma Separated Variable file and is a format that most spreadsheet programs will recognize 8 Now select the folder and enter...

Страница 18: ...n the standard width of the cell Click on the A at the top of the column and then select Format Column and Autofit Selection Or simply drag the column separator between A and B to achieve the right wi...

Страница 19: ...calibration is required at least once every 12 months to replace the electrochemical oxygen sensor Where the instrument is used as a standby for use in emergencies replace the main batteries as detail...

Страница 20: ...effects Body Part Effect First Aid Procedures Skin Contact could result in a chemical burn Wash the affected part with a lot of water and remove contaminated clothing If stinging persists get medical...

Страница 21: ...cklight times out after 15 seconds when turned on Oxygen Sensors See Appendix on page 22 Analox 9100 9212 9HSUB oxygen sensor with microprocessor applied temperature compensation 2 3 year life at 21 S...

Страница 22: ...replacement or a refund of net purchase price paid as per our Invoice on initial purchase from us We shall have no liability for losses damages costs or delays whatsoever We shall have no liability f...

Страница 23: ...Version Document Ref AS2 802 14 January 2016 Page 21 12 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Страница 24: ...yields the same partial pressure of oxygen as that present in the current environment SEV on the Sub MkIIP uses 1000mbara as its surface pressure for a point of reference For example 1 If the atmosph...

Страница 25: ......