10.3

 

Zero calibration 

 

 

NOTE: 

WHEN SETTING THE GAS FLOW RATE ON THE 
CONTROL VALVE, SO LONG AS THE FLOW 
INDICATION BALL IS RAISED OFF THE BOTTOM OF 
THE GAS FLOW INDICATOR THIS IS SUFFICIENT 
FLOW TO CALIBRATE THE INSTRUMENT ACCURATELY. 

 

 

 

Fit the control valve to the zero gas. 

 

Push the calibration valve pipe fully into the push-fit connector of the 
sensor housing of the unit. 

 

Fully open the control valve by turning the control knob anti-clockwise. 

 

Allow the gas to flow for 2 minutes. 

 

Enter Technician Mode by pressing the mode switch 3 times. If entered 
successfully the green LED will flash off for 1.5 seconds and on for 0.5 of a 
second. 

 

Select Zero Calibration by pressing the mode switch 5 times. The Red 
Alarm 1 LED will light up to show you are now in this mode. 

 

Press the mode switch 2 times to start the Zero Calibration, the Red Alarm 
1 LED will turn off and the Green LED will continue to flash. 

 

Wait one minute for the instrument to adjust. When the instrument has a 

new calibration value, the buzzer will sound one bleep and all the LED’s 

will be off. 

 

Accept this new calibration value by pressing the mode switch 2 times. 
The green LED will flash to show the instrument has accepted the new 
Carbon Monoxide value. 

 

To return to normal operation, press the mode switch once. The LED’s and 

buzzer will illuminate / sound 4 times before returning to normal 
operation. 

 

Fully close the control valve by turning the control knob clockwise. 

 

Remove the calibration valve pipe from the elbow push-fit connector of 
the sensor housing of the unit. 

 

 

NOTE: 

IF AT ANY TIME YOU WOULD LIKE TO ABORT ZERO 
CALIBRATION, PRESS THE MODE SWITCH ONCE AFTER 
STEP 9), THIS WILL BRING YOU BACK TO TECHNICIAN 
MODE WITHOUT ACCEPTING THE NEW CARBON 
MONOXIDE VALUE. PRESS THE MODE SWITCH ONCE 
AGAIN, THIS WILL BRING YOU BACK TO NORMAL 

OPERATION, THE LED’S AND BUZZER WILL 

ILLUMINATE / SOUND 4 TIMES.  ALTERNATIVELY, 
DISCONNECT THE POWER SUPPLY TO THE 

ANALOX

 CO 

CLEAR

, WAIT A MOMENT AND RE-CONNECT POWER. 

 

 

 

Содержание CO Clear

Страница 1: ...Analox Ltd 15 Ellerbeck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net CO Clear User Manual Commercial in Confidence...

Страница 2: ......

Страница 3: ...ble 10 6 1 AC supply models 10 6 2 DC supply models 10 6 3 Alarm repeaters 10 6 4 Alarm relay output models 10 6 5 Relay wiring 11 6 6 4 to 20mA output models 12 7 Operation 13 7 1 Normal operation 13...

Страница 4: ...re re use Obtain medical advice if continued irritation Ingestion Corrosive May cause sore throat abdominal pain nausea and severe burns of the mouth throat and stomach and may be fatal If swallowed D...

Страница 5: ...V AC and 230V AC Versions Less than 5W 24V DC Version Mains frequency 50 60Hz The Remote Alarm Repeater has ingress protection to IP43 direct sprays of water up to 60 from the vertical in accordance w...

Страница 6: ...r please check you have the following 1 Analox CO Clear main unit and Alarm Repeater 2 User Manual 3 Test Certificate 4 Rawl Plugs and Screws for Wall Mounting 5 Drilling Template 6 Analox CO Clear Wi...

Страница 7: ...10 250V AC supply 2 110 120V AC supply 3 9 24V DC supply The Analox CO Clear is intended to be used as a permanent installation It provides a digital readout of Carbon Monoxide and 2 audio and visual...

Страница 8: ...e CO cell increases by what we call the compensation factor see Table 1 For example if a CO Clear He measured 3ppm CO in a tank containing 99 He the actual CO content would be 2 1ppm 3 x 0 7 2 1ppm Th...

Страница 9: ...wall and use the Rawl plugs and screws provided to mount the unit It is not necessary to dismantle the Analox CO Clear main unit in any way prior to installation You need to ensure the mains plug fus...

Страница 10: ...ear enclosure An 8 metre 8 core cable is pre wired to the Analox CO Clear on units to be fitted with an alarm repeater This ensures that for a basic installation there is no need to dismantle the Anal...

Страница 11: ...ened Test that the cable is adequately gripped by the cable gland Ensure that the cable is suitable for purpose the load is within the limits of the relay 240VAC 28VDC 2Amps and the insulation of the...

Страница 12: ...20mA output models You may have ordered your Analox CO Clear with a 4 to 20mA output A 2m cable is factory fitted to the instrument The 4 to 20mA output current is generated by the instrument The cus...

Страница 13: ...n the first alarm level then the Alarm 1 indicator will begin to flash and the buzzer will sound at its slow speed If the measured concentration of CO continues to rise above the second alarm level th...

Страница 14: ...ollowing specifications Maximum inlet pressure 200Bar Outlet pressure 0 3 BarG For systems where 1 4 pipe work is used the push fitting in the sensor housing must be swapped out Remove the sensor hous...

Страница 15: ...e This housing allows the Carbon Monoxide sensor to be easily replaced when necessary To ensure continuous operation it is recommended to replace the sensor at 18 month intervals The procedure for rep...

Страница 16: ...l only connect in one orientation NOTE THE NEW CARBON MONOXIDE SENSOR IS SUPPLIED IN A HOUSING READY TO BE FITTED Carefully feed the sensor wire and sensor housing through into the bulkhead and then f...

Страница 17: ...to perform a calibration check you will need the following equipment 10 1 0 to 10ppm in nitrogen range calibration gases Air and Nitrox Kit for use with Analox CO Clear 1 Zero Gas 20 9 O2 in N2 balan...

Страница 18: ...the Zero Calibration the Red Alarm 1 LED will turn off and the Green LED will continue to flash Wait one minute for the instrument to adjust When the instrument has a new calibration value the buzzer...

Страница 19: ...t has accepted the new Carbon Monoxide value To return to normal operation press the mode switch once The LED s and buzzer will illuminate sound 4 times before returning to normal operation Fully clos...

Страница 20: ...ther mode is selected press the Mode switch once to return to Technician Mode and then repeat this selection The display will indicate the present value of the alarm Press the Mode switch twice to pro...

Страница 21: ...t the problem contact your qualified service engineer A summary of the indicator lamps and buzzer operations is shown below OK LAMP GREEN ALARM1 LAMP RED ALARM2 LAMP RED FAULT LAMP YELLOW MEANING OFF...

Страница 22: ...0x70 mm 6 8x5 11x2 75inches IP Rating Analox CO Clear IP65 Sensor Type Electro Chemical Cell Sensor Life Expected 2 years at Standard Temperature and Pressure Sensor Warranty 9 months graded Electroni...

Страница 23: ...f 24 Copyright 2006 Analox Ltd All Rights Reserved Commercial in Confidence 13 Disposal According to WEEE regulation this electronic product cannot be placed in household waste bins Please check local...

Страница 24: ...CO Clear User Manual Int Approved Document ref ACO 810 28 October 2018 Page 24 of 24 Copyright 2006 Analox Ltd All Rights Reserved Commercial in Confidence 14 Declaration of conformity...

Отзывы: