6.5
Relay wiring
The cable gland is for cables of outside diameter between 5 and 7mm, if cable fitted is
outside that range, a suitably specified cable gland must be used. Ensure that the gland is
properly tightened. Test that the cable is adequately gripped by the cable gland. Ensure
that the cable is suitable for purpose, the load is within the limits of the relay,
240VAC/28VDC, 2Amps, and the insulation of the external circuit meets the requirements
for basic insulation 240VAC/28VDC, 2 Amps. After completing wiring, ensure that the
terminal box cover is securely replaced.
Figure 1
Relay terminal box terminations
Figure 2
– Relay Connections
Once the Analox CO
Clear
installation has been completed a calibration should be
performed in accordance with section 10 of this manual.