![Analog Devices EVAL-AD7625FMCZ Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/analog-devices/eval-ad7625fmcz/eval-ad7625fmcz_user-manual_2940063013.webp)
EVAL-AD7625FMCZ/EVAL-AD7626FMCZ User Guide
UG-745
Rev. A | Page 13 of 28
DC TESTING—HISTOGRAM
Figure 21 shows the
Histogram
tab. The histogram can be used
to test the ADC for the code distribution for dc input, compute
the mean and standard deviation or transition noise of the
converter, and display the results. Raw data is captured and
passed to the PC for statistical computations.
To perform a histogram test, click the
Histogram
tab, then click
Single Capture
or
Continuous Capture
.
A histogram test can be performed without an external source
because the evaluation board has a buffered V
REF
/2 source at the
ADC input.
To test other dc values, apply a source to the J3 and J5 inputs.
The signal may need to be filtered so that the dc source noise is
compatible with that of the ADC.
AC TESTING—HISTOGRAM
The histogram can also be used to test the ADC for the code
distribution for ac input, compute the mean and standard
deviation or transition noise of the converter, and display the
results. Raw data is captured and passed to the PC for statistical
computations.
To perform a histogram test, click the
Histogram
tab, then click
Single Capture
or
Continuous Capture
.
An AC histogram requires a quality signal source applied to the
input J3 and J5 connectors.
Figure 21 shows the histogram for a 20 kHz sine wave applied to
the ADC input and the results calculated.
The
Histogram Analysis
boxes, Label 1 in Figure 21, show the
various measured values for the data captured.
Figure 21. Histogram Capture Tab
1
12552-