background image

Preliminary Technical Data 

EVAL AD5933-U1

 

Rev. PrA | Page 9 of 16 

To set the frequency sweep step size, enter 

20

 (Hz) in the 

Delta 

Frequency

 box (Arrow 1A). The frequency step size is 24-bit 

accurate. 

To set the number of increments along the sweep, enter 

511

 in 

the 

Number Increments

 box (Arrow 1A). 511 is the maximum 

number of increments that the device can sweep across. The 
value is stored in a register as a 9-bit value. 

To determine the delay between the time a frequency increment 
has taken place on the output of the internal DDS core and the 
time the ADC samples the response signal, enter 

15

 (cycles) in 

the 

Number of Settling Time cycles

 box (Arrow 1A). For 

example, if the next output frequency is 32 kHz, the delay 
between the time the DDS output this frequency and the time 
the ADC samples the response signal is 15 × (1/32 kHz). The 
maximum number of settling time cycles delay that can be 
programmed to the board is 511 cycles. The value is stored in a 
register as a 9-bit value. This value can be further multiplied by 
a factor of 2 or 4, as explained next. If you are sweeping across a 
high Q structure such as tuning fork or resonant circuit, make 
sure that the settling time is sufficient such that the structure 
settles before incrementing the frequency step. This is achieved 
by increasing the settling time.  

Set the output excitation voltage gain control (either 
2/1/0.5/0.2 V p-p) of the AD5933 (at Pin 6,VOUT1 which is 
connected to evaluation board test pin T8 to 1 V p-p (see th

 

Evaluation Board Schematic). Set the post gain of the ADC 
response stage (either X1 or X5) to 

X1

 (Arrow 1B). To choose 

the internal RC oscillator as the system clock, check 

Internal 

RC Oscillator 

in the 

System clock

 panel. NOTE: For a narrow 

bandwidth/high Q response, for example, tuning fork, make 
sure that the ADC gain is set to a gain of 5 and the output 
voltage is set to 1 V p-p.  

Choose a series resistor or a resistor in parallel with a capacitor 
in the 

Calibration Impedance

 panel (Arrow 1B). For this 

example, choose 

capacitor only

 to measure the impedance 

sweep across a capacitor. For this example, enter 

0

 (Ω) the 

Resistor Value

 box, and 

15E-12

in the 

Capacitor Value

 box.  

To program the sweep conditions register values into the 
AD5933 RAM through the I

2

C interface, click 

Program 

Registers

 (Arrow 2). As explained in previously, the value 

programmed into the settling time cycles can be further 
multiplied by a factor of 2 or 4 for a sweep (Arrow 3A). Click 

X1

 (the default), 

X2

, or 

X4

. To place the AD5933 into standby 

or power-down mode, click 

Enter Standby Mode

 or 

Enter 

Power Down Mode

.  

Click

 Calculate Gain Factor

 (Arrow 3B). An explanation of the 

gain factor, which is the admittance per code of the calibrated 
system, is provided in detail in the product data sheet. This 
results in the gain factor being calculated for the calibration 
impedance, which is returned to the interface and subsequently 
used for the sweep across either the impedance under test. If 
you want to use a gain factor that has been previously 
calculated, click the returned gain factor and enter the new gain 
factor.  

05435-020

 

Figure 14. 

After the system calculates the gain factor for the current 
system setup, it appears in the 

calculated gain factor 

box. 

To re-use the sweep setup under conditions other than those 
that were initially programmed, or to reprogram different sweep 
conditions, recalculate a new gain factor and re-enter it into the 
interface to read back any significant results. The gain factor 
calculated in software is not programmed into the AD5933 
RAM and is valid only when the software program is running. 
It is not retained when the software program is closed. However, 
it is retained when the part is placed into power-down mode. 

To begin the sweep, click

 Start Sweep

 (Arrow 4). This returns 

a plot of the impedance vs. frequency for the impedance 
(Figure 13).The progress of the sweep is outlined with a 
progress bar (Figure 15).  

05435-028

 

Figure 15. 

A message indicates that the sweep is complete. 

To take a reading from the on board temperature sensor, click 

Read Temperature

. This returns the 13-bit temperature of the 

device. See product data sheet for more information on the 
temperature sensor. 

Содержание EVAL AD5933-U1

Страница 1: ...mbines an on board frequency generator with a 12 bit 1 MSPS ADC The frequency generator allows an unknown external impedance to be excited with a known frequency The response signal from the impedance...

Страница 2: ...uration and Function Descriptions 5 Getting Started 6 Setup Sequence Summary 6 Step 1 Install the Software 6 Step 2 Connect the USB Cable 7 Step 3 Verify the Links and Power Up The Evaluation Board 7...

Страница 3: ...AD5933 Analog Circuitry Input Supply AVDD REF is decoupled to the analog ground plane AGND J9 2 via standard 0 1 F and 10 F suppression capacitors J10 2 AGND Terminal Block Analog Ground Connection De...

Страница 4: ...an operational amplifier LK7 This link is inserted to connect the Tf2 OUT SMB connector to the inverting terminal of the transimpedance amplifier within the AD5933 This link is required to complete th...

Страница 5: ...edance amplifier Test impedance is connected between this pin and the VOUT pin 6 VOUT Output AC Excitation Signal Programmable frequency range 0 Hz to 50 kHz Test impedance is connected between this p...

Страница 6: ...aluation board software to run the sweep function See Step 4 Perform a Frequency Sweep STEP 1 INSTALL THE SOFTWARE Place the CD accompanying the evaluation board into the CD drive of the user computer...

Страница 7: ...not passed the Windows logo testing to verify compatibility with Windows XP This error appears because this is an evaluation setup installation and not meant to be used in a production environment Cli...

Страница 8: ...ace program is run it appears as shown in Figure 13 The figure shows the interface along with an impedance profile for a sweep across 15 pF ceramic capacitive impedance This section describes how to s...

Страница 9: ...this example choose capacitor only to measure the impedance sweep across a capacitor For this example enter 0 the Resistor Value box and 15E 12in the Capacitor Value box To program the sweep condition...

Страница 10: ...the impedance of the network under analysis varies across the programmed frequency range To view how the relative phase across the network under analysis varies click relative phase vs frequency The...

Страница 11: ...e Device Manager Expand Other devices Figure 22 05435 017 1 EXPAND THIS DIRECTORY 2 RIGHT CLICK ON THIS DEVICE Figure 22 The computer has not recognized the USB device that is the AD5933 evaluation bo...

Страница 12: ...15 14 13 12 11 10 9 T1 T2 T5 T6 R2 50 LK7 OP97 U1 OUT 6 7 4 2 3 VIN TRIM VOUT GND U4 ADR425 C32 0 1 F C31 10 F 2 5 6 4 CLK2 LK6 LK2 LK3 LK4 LK5 C25 10 F C26 0 1 F C30 10 F C28 0 1 F C2 10 F C24 0 1 F...

Страница 13: ...OE PA3 WU2 PA4 FIFOADR0 PA5 FIFOADR1 PA6 PKTEND PA7 FLO SLCS RDY0 SLRD RDY1 SLWR IFCLK RSVD AGND GND GND GND GND GND GND GND AVCC VCC VCC VCC VCC VCC VCC VCC VCC WP SCL SDA U2 24LC64 C22 0 1 F 3 3V 8...

Страница 14: ...8 ADP3300 3 3 8 7 1 IN1 2 6 3 SD GND OUT2 ERROR NR IN2 5 1 4 C16 0 1 F T1 T2 T3 T4 C15 0 1 F C14 10 F 3 3V D4 3 3V R24 1k Figure 26 05435 027 C12 10 F C11 0 1 F DVDD_5V DGND J2 2 J2 1 C35 10 F C36 0 1...

Страница 15: ...ve device Electrostatic charges as high as 4000 V readily accumulate on the human body and test equipment and can discharge without detection Although this product features proprietary ESD protection...

Страница 16: ...5933 U1 Preliminary Technical Data Rev PrA Page 16 of 16 NOTES 2005 Analog Devices Inc All rights reserved Trademarks and registered trademarks are the property of their respective owners EB05435 0 3...

Отзывы: