ADMV8913-EVALZ
Evaluation Board User Guide
UG-1952
Rev. 0 | Page 9 of 17
PERFORMING EVALUATION
ADMV8913-EVALZ QUICK START
To set up the ADMV8913-EVALZ, take the following steps:
1.
Connect the RF1 and RF2 ports to a network analyzer (or a
similar instrument). Typically, RF1 and RF2 are connected
to Port 1 and Port 2 on the network analyzer, as shown in
Figure 2.
2.
Connect the
SDP-S
to the 120-pin connector on the
ADMV8913-EVALZ. Do not connect the SDP-S to the PC
until after completing Step 4 or Step 5.
3.
On the ADMV8913-EVALZ, slide the S2 switch down (as
shown in Figure 2) to select the SPI.
4.
On the ADMV8913-EVALZ, slide the S1 switch down (as
shown in Figure 2) to power the ADMV8913-EVALZ from
the 5 V USB supply voltage from the PC via the SDP-S.
5.
Alternatively (to Step 3), slide the S1 switch up and connect
power supplies to the VPOS and VNEG ports. The applicable
voltage range for VPOS is between 3.5 V and 5.5 V, and the
applicable voltage range for VNEG is between −5.5 V and
−2.7 V. The external supply current limits must be set to
20 mA. The expected supply current drawn for VPOS is
12 mA to 14 mA, and for The expected supply current
draw for VNEG is 2 mA to 3 mA. The
ADMV8913
chip
current drawn per supply pin is typically 10s of microamps
or less. Most of the current drawn from the ADMV8913-
EVALZ comes from the LDO regulators and the status
indicator light emitting diodes (LEDs), DS1 to DS3.
6.
Connect a USB cable between the PC and the SDP-S.
7.
Open the
ACE
software. The
ADMV8913 Board
appears
in the
Attached Hardware
section of the
Start
tab. Double
click the
ADMV8913 Board
plug-in to open two tabs, one
for the
ADMV8913 Board
plug-in view and one for the
ADMV8913
chip plug-in.
8.
Use the
CONFIGURATION
section (see Figure 9) in the
ACE software to initialize the chip. By default, the
ADMV8913_Register_Load_1.csv
file is loaded into this
section. Click
Apply
to send the default settings to the chip
and to allow the main diagram user controls to become
editable.
26169-
009
Figure 9. ADMV8913 CONFIGURATION Section
NETWORK ANALYZER SETTINGS
When evaluating the ADMV8913-EVALZ, a good starting point
for configuring the network analyzer is as follows:
•
Start frequency = 0.1 GHz
•
Stop frequency = 40 GHz
•
Number of points = 400
•
Step size = 100 MHz
•
Power level = −10 dBm
•
Measure types = S-parameters (S21, S11, and S22)
•
Format = log magnitude
•
Calibration = full 2-port