NanEye
NanEye Viewer Software
Eval Kit Manual
• PUBLIC
UG001036
• v3-00 • 2022-May-31
43
│ 39
3.4.2
NanEye Viewer Microsoft .NET Framework Error
Figure 56:
NanEye Viewer Framework Error Message
This error occurs when libraries are missing. Therefore, to load them:
●
Go to C:\Program Files (x86)\ams OSRAM\NanEye Viewer vx.x.x.x\libraries
●
Run vcredist_x86_15.exe
3.4.3
Error Message “Can’t read from bulk pipe. System Error: 0x00000079/ Not
receiving images from the sensor
”
●
This indicates that the communication between sensor and EVK may have problems. Therefore:
●
Check the sensor cable wires condition and its connector soldering
●
If the sensor wires on connector are loose, if possible, please solder it
(1)
Check NanEye2D Module Handling Manual (NanEye_AN000493)
●
If the wires are disconnected near the sensor head zone, no repair is possible
●
Check EVK connectors status (visual inspection)
3.4.4
Error Message “There is no NanoUSB connected / Camera not connect”
●
Go to NanoUSB2 section 2.2.2
3.4.5
Error Message “External component has thrown an exception”
●
Check sensor cable wires condition and its connector soldering
●
Go to NanoUSB2 section 2.2.2
●
Go to FOB2.0 section 2.1.3