the AC adapter must be connected for 4 hours in 
order to fully restore the charge. If you need a 
replacement AC/DC Battery Charger for the 630A, 
contact the AMPTEC customer service department 
and request an option “630DC” Battery Charger. 

A-5.

Setup and Use 

Once the AMPTEC 630ES has had it’s batteries 
charged for 12 to 24 hours it is ready for use. The 
630ES consumes little power and generates virtually 
no heat.  Consequently, it may be used in any area 
where the environment does not exceed the 
specifications of Table B-2.

Avoid exposing the 630ES Explosive Safety Igniter 
Tester to extremes of temperature which will affect 
accuracy and shorten battery life-span. 

630ES Explosive Safety Igniter Tester keyed 
connector. In this way the user may not ever be able 
to connect any test leads while the unit is charging. 
The 630ES is designed so it is virtually impossible to 
be powered (in operating - measurement mode) 
directly from the AC line adapter. As an additional 
note, the 630ES main power switch 

must also be

 in 

the “

Off/Charging

” 

mode

 in order for the connected 

battery charger to recharge the 630ES’s batteries.    

For customers using 220VAC 50 HZ AC line power 
for running the battery charger, please inform 
AMPTEC’s sales  department at the time of your 
order and the appropriate adaptor will be included 
with the 630ES Failsafe Ohmmeter shipment for an 
additional fee.

A fully charged battery pack may power the 630ES 
for approximately 8 to 10 hours before requiring a 
recharge. Amptec installs a quality set of 4 each 
Heavy Duty (5700 mAHr) Ni-Cad batteries. The 
620A will also operate on a 4000 mAhr D cell Ni-
Cads with a shortened operating time between 
charges.

The “Power” switch has two separate modes. The 
“ON” position supplies internal battery isolated 
power to operate the 630ES when they are charged. 
If you turn on the 630ES, and the display does not 
come on, it may indicate the batteries need charging. 
The “OFF/CHARGING” power switch position is for 
use when the batteries need charging or the 630ES is 
not in use. As mentioned earlier of course the AC/DC 
Battery Charger must be plugged into the 630ES’s 
keyed connector to facilitate charging the batteries.  

Although the batteries are fully charged prior to 
shipment, it may be desirable to refresh the charge 
for 24 hours before use.  As a rule of thumb, the 
630ES requires twice as much time to fully recharge 
as the amount of discharge time.  For example, if the 
instrument was used continuously for 2 hours,

Содержание 630ES

Страница 1: ...RESEARCH AMPTEC PHONE 512 858 4045 800 350 5105 website http www amptec com FAX 512 858 4340 TOLL FREE IN USA Revision D AMPTEC 630ES SAFETY IGNITER TESTER OPERATION MAINTENANCE MANUAL...

Страница 2: ...ry Charger Power Requirements A 5 Setup and Use SECTION B 630ES Explosive Safety Igniter Tester SPECIFICATIONS Table B 2 Specifications SECTION C REPLACEMENT OPTIONAL AND ACCESSORY ITEMS C 1 Available...

Страница 3: ...y Scheme Figure E 1 Model 630ES Block Diagram Figure E 1B Analog to Digital Converter Timing Diagram Figure E 2 Analog Section of IC1 and IC2 Figure E 3 Constant Diagram Simplified Figure E 4 LED Disp...

Страница 4: ...it useful in a variety of applications Please check the front of this manual for any addendums that may apply to new 630ESs and 630AN conversions SECTION A UNPACKING AND INSPECTION A standard feature...

Страница 5: ...630ES can be opened If the Purge Valve was opened return the valve to the closed state tighten clockwise once the 630ES lid is open The 630ES has a way to remove it s lid completely for example when u...

Страница 6: ...ng 220VAC 50 HZ AC line power for running the battery charger please inform AMPTEC s sales department at the time of your order and the appropriate adaptor will be included with the 630ES Failsafe Ohm...

Страница 7: ...ge clamped at 1 6 volts 2 Power 4 D 5 0 AHr Heavy Duty 1 2V rechargeable nickel cadmium batteries Fuse 2 ampere fast blo 3 AG type AC DC Battery Charger Option 630DC provides 9VDC at 1 1 Ampere nomina...

Страница 8: ...d set is terminated with the mating connector for 630ES Front panel connection port user specified AMPTEC RESEARCH s website http www amptec com should also provide the latest information on accessori...

Страница 9: ...nd phone changes information C 1 Available Accessories and Options This manual does not list all possible accessories that AMPTEC RESEARCH is willing to provide as a support items for the 630BN Ignite...

Страница 10: ...tip Option 630 304 Gold 4 Wire Kelvin Clip Test Lead Set Option KCS Kelvin Clips accept banana jacks in back works with Option 630 403 is a 4 wire Kelvin Handheld Probe Lead set terminated with two ha...

Страница 11: ...or the 20 Ohm lowest resistance range is just below the display on the left see item 20 of the 630ES Front Panel Diagram When a given range is selected an indicator LED informs the user Also note that...

Страница 12: ...ion AMPTEC 630ES Front Panel Diagram 1 Calibration Maintenance access screws also can be easily protected with a tamper proof calibration sticker or seal 2 Purge Valve provides air pressure equalizati...

Страница 13: ...13 10 ohm test resistor for checking the mid scale functionality of the 20 ohm range 14 Input jack for connecting to the 630ES Ohmmeter common test lead plug in black banana jack This common jack is a...

Страница 14: ...stepped down from the 9VDC 1 1A supplied by the charger The correct charging voltage is supplied by the adapter included with the instrument Additional AC DC Battery Chargers are available as Option 6...

Страница 15: ...keyed test lead connector is optionally available for customized wiring connections kelvin clips cables terminated with spade lugs and special banana jacks All AMPTEC ohmmeters use a high impedance v...

Страница 16: ...ng the test resistors in the Functional Test Section then the connect ion problem must be outside of the 630 series meter i e your wiring harness or the actual device under test connection If the 630...

Страница 17: ...the fuse from the fuseholder With an adjustable DC power supply set the power supply output to be 4 50 VDC Be sure to observe power supply polarity Connect the power supply to the test points labeled...

Страница 18: ...te the batteries need charging or fuse needs replacing If the 630ES exhibits problems that cannot be eliminated by reviewing Chapters B and D the following guidelines have been established to help sol...

Страница 19: ...or move the replacement part until the solder has cooled Cold solder and bad solder joints can cause more problems Use the chassis ground connect to the common terminal of the functional test section...

Страница 20: ...03 takes care of the digital part of the 4 digit 20 000 count dual slope conversion COUNTERS DISPLAY COMPARATOR BUFFER MULTIPLEXER ZERO CROSSING DETECTOR INTEG UUT CONTROL LOGIC INPUT SWITCH DC CONSTA...

Страница 21: ...EGRATOR INTEGRATOR COMPARATOR COMPARATOR CAZ CAZ CSTRAY CSTRAY RINT RINT FIGURE E 2B PHASE ll INTEGRATE INPUT FIGURE E 2C PHASE lll AND DE INTEGRATE CINT CINT VIN VIN CREF CREF 3 3 1 1 2 2 6 6 5 5 4 4...

Страница 22: ...ished in Phase I The time or number of counts required to do this is proportional to the input voltage E 4 1 Reference Voltage The precision reference voltage required to do the A D conversion is deve...

Страница 23: ...connect the 5 volt supply directly across R218 D202 the range resistor and R L D203 however acts as a 1 6 volt zener diode limiting the voltage that can appear across these components Even if every co...

Страница 24: ...Power Supply Scheme The 5 volt power supply is provided directly by the batteries for driving the LED displays and digital logic The 5VD is used for driving IC8 the low battery detection circuit The 1...

Страница 25: ...um of E plus the output of in IC202 Consider now that the short is removed from the Ihi and I terminals and a 100 ohm resistor R is lo L connected in its place The current through RL increases the vol...

Страница 26: ...hm range RL2 is for the 200 Ohm range RL3 is for the 2K Ohm range and RL4 is for the 200K Ohm range A B G F C E D D P SEGMENT PIN A 1 B 13 C 10 D 8 E 7 F 2 G 11 D P 6 Common 14 PIN 1 2 3 4 5 6 7 PIN 1...

Страница 27: ...g the calibration procedure Some calibration adjustments are accessed by removing the calibration access screws cont next column in the top plate perimeter of the unit Use the two handles careful not...

Страница 28: ...2 Ohm Range Full Scale Adjustment 1 Select the 2 ohm range Connect the 4 Wire Kelvin leads to the 1 00 ohm standard resistor 2 Adjust RV6 on main PCB for a display reading of 1 0000 Ohms 3 Check the 2...

Отзывы: