background image

AMPTEC 620VN

620VN

620VN-501

620VL

T

-602

620VN

 OP 24/7 DC

620VN-123

D-3 

RS232C Computer Control and Command Set      

750 mA

620VN  Rear Panel with options RS232C I/O  

The rear panel

 (shown above) 

may contain optional 

jacks, terminals,  labels, and stickers. 

Item 1

 -The option "247" Continuous Isolated 

Operating Power DC Adapter (item #1) must be left 
in the rear panel jack in order to continuously operate 
the unit. 

For Safety Reasons 

Only the AMPTEC 

supplied AC/DC adapter, in conjunction with the 
meter’s OP 247 internal DC/DC circuitry has been 
tested and approved for its 3 KV isolation.

  

Item 2

 is the fuse holder - replace with a 2 ampere 3 

AG type fast blow fuse (rarely needs replacing). 

RS232C Serial Interface - Item # 3 is the safety 
isolated RS232C serial I/O (9 pin D type Sub-min 
connector ). This RS232C Interface provides the 
meter’s resistance measurements to a PLC fitted with 
a serial port. The 

RS232C protocal settings should 

be 9600 Baud, 8 Bits , No parity , 1 Stop Bit ,

 9 pin 

D Sub-min connection

RS232C Command Set  (Option 232 )

(NOTE: All front panel range buttons must be de-
selected (press in half way) 

with all range 

pushbuttons out

 in order to disable with local 

lockout and enable RS232C control. 

Note commands 

are case sensitive

.

C

Continuous Read Mode  - 

RS232C I/O 

 

outputs a data string every A to D conversion cycle, 
approx

 2.5 times per second.

S

Single Read Mode - 

RS232C I/O 

 outputs 

a single data string upon reception of a “R” 
command.

R

Read - 

Commands 

RS232C I/O 

 to output a 

single data string

 

( 1 resistance reading).

r

0

De-Selects all Ranges 

r1

Selects the 20 Ohm Range

r2

Selects the 200 Ohm Range

r3

Selects the 2000 Ohm Range

r4

Selects the 20 K Ohm Range

 

r5        

Selects the 200 K Ohm Range

  

r6        

Selects the 2 Meg Ohm Range

 

V

Version

 c

ommands board to output the firmware 

version string.

Data Format - 

The 

RS232C I/O

 outputs a data string 

with the following format:

1.2345E+3 

  

The measurement is always in Ohms

 

+3

(where E+3 = 10

scientific notation style)

 

.  The 

Exponent is defined below.

 1.2345E+3 = 1.2345 

+3

+3

KOhms 

(where E

=10

). 

1.3700E+1

 = 

13.700 Ohms

 

+1

+1

(where E

=10

Range

  Exponent

 

20.0   Ohm

  E+1

    

200.0 Ohm

  E+2    

                        

                

   

    

  

2000  Ohm

  E+3     

20   KOhm

  E+4     

200 KOhm          E+5    
2.0   M Ohm        E+6  

An overrange condition is indicated by                     
9.9999E

nn..

Where nn is the

 

selected resistance range

 

exponent.  A Range Error is indicated by x.xxxxERR.

 

Item # 4

 is the unit’s serial number sticker.  

Item # 5

 is the 4 terminal rear terminal strip (they are 

wired in parallel with the front terminals). If a “2 wire 
ohms” connection is made then the 

V

 high and 

high 

terminals should be shorted together, and the 

V

 low and 

I

 low terminals should be shorted. The 620VN Voltage 

High, Voltage low, Current high and Current low wires 
are permanently connected to the gold plated rear termi-
nal strip ( seel labeled terminal strip on rear panel). 

 
Item #6  Calibration Sticker - 

If the calibration due 

date has expired ( 1 year) AMPTEC or a Cal. Lab can 
contacted to re-certify the AMPTEC 620VN Explosive 
Safety Ohmmeter/Igniter Tester. contact AMPTEC 
customer service 001-512-858-4045 .  

www.amptec.com

page 8

Содержание 620VN

Страница 1: ...OPERATION MAINTENANCE MANUAL AUSTIN TEXAS 78759 USA CAGE CODE 1CRL2 PHONE 001 512 858 4045 or 1 800 350 5105 from the USA FAX 001 512 3858 4340 or 1 800 430 5440 from the USA website http www amptec...

Страница 2: ...Permission and a return authorization RMA number must be obtained directly from AMPTEC s customer service department via phone FAX or email for repairs warranty or otherwise We need to issue you an RM...

Страница 3: ...B 1 Specifications SECTION C REPLACEMENT OPTIONAL ANDACCESSORYITEMS C 1 Available Accessories and Options SECTION D OPERATION FUNCTION AND USE D 1 General D 2 Front Panel D 3 RS232C Computer Control a...

Страница 4: ...ctrical Schematics Rev E Main PCB 620RMS 070 DVM1 1of 2 and 620VL 070 Current I source 2 of 2 Source 620VN Main PCB Bill Of Materials components parts list page 1 of 2 620VN Main PCB Bill Of Materials...

Страница 5: ...n power as a floating 5VDC A 3000 VAC Isolation Hipot Test Report is available from AMPTEC s engineering department that has established the isolation of the 2 stage OP247 as being at least Isolated t...

Страница 6: ...er Remote Control Interface The AMPTEC 620VN safety meter is normally configured with the option 232 an optically Isolated Serial Interface It that allows computer controller to remotely send I O comm...

Страница 7: ...ce of the device under test Option 247 for continuous operation 24Hr 7day Dual stage isolation provides safe power for 620 series Igniter Tester to safely run continuous 24 Hrs day without batteries I...

Страница 8: ...link which indicates overrange whichis also available via the automated control RS232C Interface D 1 General Operation This section contains operating instructions for the AMPTEC 620VN Explosive Safet...

Страница 9: ...eception of a R command R Read Commands RS232C I O to output a single data string 1 resistance reading r0 De Selects all Ranges r1 Selects the 20 Ohm Range r2 Selects the 200 Ohm Range r3 Selects the...

Страница 10: ...l 4 Ensure that communication specifications are as follows 9600 bits per second 8 data bits No parity bit One 1 stop bit No flow control 5 Start testing each range of the unit via the RS232C interfac...

Страница 11: ...ing displayed on screen corresponding to the range that the unit is currently operating in ii Must be in Single Read Mode to receive this command properly C i Should receive prompt Continuous Read Mod...

Страница 12: ...range selected on the AMPTEC 620 tester Connecting VHI to IHI and Vlo to Ilo eliminates the wandering open circuit display condition Figure D 1 above illustrates the 4 wire princi ple eliminates lead...

Страница 13: ...urement error if not compensated for Many Ordinance test procedures have the 620 series Igniter Tester user short their wiring harnesses at the very end by the squib and record the resistance value or...

Страница 14: ...Igniter Tester and the display does not come on it usually means the AC to DC Adapter isn t supplying power or fuse needs replacing If the 620VN exhibits problems that cannot be eliminated by reviewi...

Страница 15: ...Do not touch or move the replacement part until the solder has cooled Cold solder and bad solder joints can cause more problems Use the chassis ground connect to the common terminal of the functional...

Страница 16: ...ICL71C03 takes care of the digital part of the 4 digit 20 000 count dual slope conversion COUNTERS DISPLAY COMPARATOR BUFFER MULTIPLEXER ZERO CROSSING DETECTOR INTEG UUT CONTROL LOGIC INPUT SWITCH DC...

Страница 17: ...ER 1 F 1 F INTEGRATOR INTEGRATOR COMPARATOR COMPARATOR CAZ CAZ CSTRAY CSTRAY RINT RINT FIGURE E 2B PHASE ll INTEGRATE INPUT FIGURE E 2C PHASE lll AND DE INTEGRATE CINT CINT VIN VIN CREF CREF 3 3 1 1 2...

Страница 18: ...g point established in Phase I The time or number of counts required to do this is proportional to the input voltage E 4 1 Reference Voltage The precision reference voltage required to do the A D conv...

Страница 19: ...ts The voltage drop across the range resistor is 0 5 volt just as it was when the output terminals were shorted The current through RL is 5 milliamperes just as it was through the jumper when the outp...

Страница 20: ...5 6 7 PIN 14 13 12 11 10 9 8 Figure E4 620 series LED Display Pin Out Detail Functions The 620VN Tester measurement circuitry is also failsafe current limited even under worst case component failure...

Страница 21: ...Option 300 Kelvin Clip Lead Set F 3 Calibration Procedure The 620VN Igniter Tester must be calibrated with a fully charged battery pack and should be allowed to warm up for a minimum of 15 minutes be...

Страница 22: ...RS232C serial interface If the low battery LED is illuminated 620VN Series Igniter Tester readings should not be trusted The AMPTEC 620VN is powered by an external battery if the meter indicates LOW B...

Отзывы: