SECTION D - OPERATION, FUNCTIONAL 
SELF - TEST AND USE

 

The AMPTEC 620VN ohms ranges can be 
computer controller selected via the RS232C 
interface (see rear panel). 

Gold Plated Five way
 Input Jacks

The AMPTEC 620 front panel input jacks are a 
gold plated variety that readily accept a number 
of user connection modalities.  The unit's 
banana post input jacks may be unscrewed 
several turns to exposed a horizontal hole thru 
the gold plated post. The AMPTEC 620 user can 
directly connect a bare wire thru the banana post 
(horizontal hole). The AMPTEC 620 front panel 
input posts also accept wires with spade lugs, 
and or banana jacks. The front panel or rear 
panel terminal strip inputs are wire for Voltage 
High and Voltage low , and Current high and 
Current low (resistance measurements). The 
spacing of the front panel jacks only allow 
horizontal connection of the dual banana jack 
test leads to prevent mis-connection (i.e. during 
calibration) .

Calibration Access Screws

Recessed in the feet on the bottom side of the  
AMPTEC 620VN meter there are 4 screws 
(phillips type). When removed you can "remove 
the lid" and access the main internal electronics 
/PCB and calibration “adjustment potentiom-
eters. They (

feet screws

) are normally meant to 

remain intact and to be removed by authorized 
Repair and Calibration Lab staff.

If the resistance being measured (including 
"Open Circuit/ Disconnected states") is a higher 
value than the selected range, the instrument's 
display will flash (blink) , which indicates 
"overrange" . whichis also available via the 

automated control 

RS232C Interface.  

D-1.

General Operation

This section contains operating instructions for 
the AMPTEC 620VN Explosive Safety Igniter 
Tester.  

D-2.  Front Panel Features and Operation

When the front panel power switch
is in the “OFF”  position, all power
is removed from the unit’s main 
PCB and the unit’s RS232C interface. When the 
620VN main power switch is placed in the 
power “ON” position, the unit powers up all 
ohms and related measurement circuitry and 
allows the optically RS232C interface located 
on the unit’s rear panel to also operate remotely. 

Resistance Range Switches

    

The 4 ½ digit display can display up to 19.999 
ohms on the 20 ohm range and 199.99 ohms on 
the 200 ohm range. The 2.0 Megohm range can 
display resistance readings up to 1.9999 
Megohms.  The meter can display from readings 
0000 to 19,999 (decimal point position varies 
with the selected range). That is referred to as 
20,000 counts, which is common for 4 ½ dight 
meters. The user can depress (feel a “click”) one 
of the meter’s range latching switches to select 
the desired resistance range. The prior range 
switch button will “pop out” when a new range 
is selected. 
The far right front panel switch labeled 
“REMOTE RS232C” must be selected (clicked 
in) to get the meter to accept remote commands 
via the RS232C (rear panel) serial interface. 
When the  “REMOTE RS232C” mode is 
selected all the meters manually selected 
resistance range switches should “pop out” or 
be de-selected. In this way, There is no manual 
selected range and meter is ready to accept 
remote RS232C computer commands for range 
and reading control.

OFF

page 7

Содержание 620VN

Страница 1: ...OPERATION MAINTENANCE MANUAL AUSTIN TEXAS 78759 USA CAGE CODE 1CRL2 PHONE 001 512 858 4045 or 1 800 350 5105 from the USA FAX 001 512 3858 4340 or 1 800 430 5440 from the USA website http www amptec...

Страница 2: ...Permission and a return authorization RMA number must be obtained directly from AMPTEC s customer service department via phone FAX or email for repairs warranty or otherwise We need to issue you an RM...

Страница 3: ...B 1 Specifications SECTION C REPLACEMENT OPTIONAL ANDACCESSORYITEMS C 1 Available Accessories and Options SECTION D OPERATION FUNCTION AND USE D 1 General D 2 Front Panel D 3 RS232C Computer Control a...

Страница 4: ...ctrical Schematics Rev E Main PCB 620RMS 070 DVM1 1of 2 and 620VL 070 Current I source 2 of 2 Source 620VN Main PCB Bill Of Materials components parts list page 1 of 2 620VN Main PCB Bill Of Materials...

Страница 5: ...n power as a floating 5VDC A 3000 VAC Isolation Hipot Test Report is available from AMPTEC s engineering department that has established the isolation of the 2 stage OP247 as being at least Isolated t...

Страница 6: ...er Remote Control Interface The AMPTEC 620VN safety meter is normally configured with the option 232 an optically Isolated Serial Interface It that allows computer controller to remotely send I O comm...

Страница 7: ...ce of the device under test Option 247 for continuous operation 24Hr 7day Dual stage isolation provides safe power for 620 series Igniter Tester to safely run continuous 24 Hrs day without batteries I...

Страница 8: ...link which indicates overrange whichis also available via the automated control RS232C Interface D 1 General Operation This section contains operating instructions for the AMPTEC 620VN Explosive Safet...

Страница 9: ...eception of a R command R Read Commands RS232C I O to output a single data string 1 resistance reading r0 De Selects all Ranges r1 Selects the 20 Ohm Range r2 Selects the 200 Ohm Range r3 Selects the...

Страница 10: ...l 4 Ensure that communication specifications are as follows 9600 bits per second 8 data bits No parity bit One 1 stop bit No flow control 5 Start testing each range of the unit via the RS232C interfac...

Страница 11: ...ing displayed on screen corresponding to the range that the unit is currently operating in ii Must be in Single Read Mode to receive this command properly C i Should receive prompt Continuous Read Mod...

Страница 12: ...range selected on the AMPTEC 620 tester Connecting VHI to IHI and Vlo to Ilo eliminates the wandering open circuit display condition Figure D 1 above illustrates the 4 wire princi ple eliminates lead...

Страница 13: ...urement error if not compensated for Many Ordinance test procedures have the 620 series Igniter Tester user short their wiring harnesses at the very end by the squib and record the resistance value or...

Страница 14: ...Igniter Tester and the display does not come on it usually means the AC to DC Adapter isn t supplying power or fuse needs replacing If the 620VN exhibits problems that cannot be eliminated by reviewi...

Страница 15: ...Do not touch or move the replacement part until the solder has cooled Cold solder and bad solder joints can cause more problems Use the chassis ground connect to the common terminal of the functional...

Страница 16: ...ICL71C03 takes care of the digital part of the 4 digit 20 000 count dual slope conversion COUNTERS DISPLAY COMPARATOR BUFFER MULTIPLEXER ZERO CROSSING DETECTOR INTEG UUT CONTROL LOGIC INPUT SWITCH DC...

Страница 17: ...ER 1 F 1 F INTEGRATOR INTEGRATOR COMPARATOR COMPARATOR CAZ CAZ CSTRAY CSTRAY RINT RINT FIGURE E 2B PHASE ll INTEGRATE INPUT FIGURE E 2C PHASE lll AND DE INTEGRATE CINT CINT VIN VIN CREF CREF 3 3 1 1 2...

Страница 18: ...g point established in Phase I The time or number of counts required to do this is proportional to the input voltage E 4 1 Reference Voltage The precision reference voltage required to do the A D conv...

Страница 19: ...ts The voltage drop across the range resistor is 0 5 volt just as it was when the output terminals were shorted The current through RL is 5 milliamperes just as it was through the jumper when the outp...

Страница 20: ...5 6 7 PIN 14 13 12 11 10 9 8 Figure E4 620 series LED Display Pin Out Detail Functions The 620VN Tester measurement circuitry is also failsafe current limited even under worst case component failure...

Страница 21: ...Option 300 Kelvin Clip Lead Set F 3 Calibration Procedure The 620VN Igniter Tester must be calibrated with a fully charged battery pack and should be allowed to warm up for a minimum of 15 minutes be...

Страница 22: ...RS232C serial interface If the low battery LED is illuminated 620VN Series Igniter Tester readings should not be trusted The AMPTEC 620VN is powered by an external battery if the meter indicates LOW B...

Отзывы: