background image

For safety reasons

, the battery charger will only 

recharge the units NICAD batteries when the  
charger is plugged into the charging jack

 and 

the 620A-4R main power switch is in the “off” 
position.

  

In this way the 620A-4R user can not 

ever make a measurement while the unit is 
charging. 

  The 620A 4R is power “On/Off” 

switch design makes it  impossible to be 
powered (in operating - measurement mode) 
directly from the AC line adapter. As an 
additional note, the 620A 4R main power switch 

must also be

 in the “

Off/Charging

” 

mode

 in 

order for the connected battery charger to 
recharge the 620A 4R’s batteries.    

A fully charged battery pack typically powers 
the 620A 4R for approximately 8  hours before 
requiring a recharge. AMPTEC installs a quality 
set of 4 each Heavy Duty (5700 mAHr) Ni-Cad 
batteries. The 620A 4R will also operate on a 
4000 mAhr D cell Ni-Cads with a shortened 
operating time between charges. Recharge time 
is typically twice the “Power On” time. An 
“Overnight” charge usually restores the 620A 
4R to a “Fully Charged” ready to use state.

The “Power” switch has two separate modes. 
The “ON” position supplies internal battery 
isolated power to operate the 620A 4R when 
they are charged. If you turn on the 620A 4R, 
and the display does not come on, it may 
indicate the batteries need charging. The “OFF/ 
CHARGING” power switch position is for use 
when the batteries need charging or the 620A 
4R is not in use.   As mentioned earlier of 
course the AC/DC Battery Charger must be 
plugged into the 620A 4R’s notched  connector 
to facilitate charging the batteries.  

As mentioned earlier of course the AC/DC 
Battery Charger must be plugged into the 620A 
4R’s notched connector to facilitate charging the 
batteries. 

Although the batteries are fully charged prior to 
shipment, it may be desirable to refresh the 
charge for 24 hours before use.  As a rule of 
thumb, the 620A 4R requires twice as much 
time to fully recharge as the amount of 
discharge time.  For example, if the instrument 
was used continuously for 2 hours, the AC 
adapter must be connected for 4 hours in order 
to fully restore the charge. If you need a 
replacement AC/DC Battery Charger for the 
630, contact the AMPTEC customer service 
department and request an option “630DC” 
Battery Charger.

When the 620A 4R is first turned on, the unit 
draws more internal power to heat up the unit’s 
ovenized zener voltage reference. If the “Low 
Battery” Indicator only comes on for a few (i.e 
10 to 15 seconds) seconds when the 620A 4R is 
first turned on then goes out, the battery levels 
are starting to indicate the charge level is low. 

A-5.

Setup and Use 

The AMPTEC 620A 4R Igniter Tester may be 
setup to operate without the unit’s lid intact. By 
removing the hinge pins the lid may be removed 
(i.e. for indoor lab bench use). Typically a test 
procedure will have the 620A 4R user test the 
620A 4R  with the units  “Functional Test 
Section” to determine it is functioning properly 
and ready for use.

Once the AMPTEC 620A 4R has had it’s 
batteries charged for 12 to 24 hours it is ready 
for use. The 620A 4R consumes little power and 
generates virtually no heat.  Consequently, it 
may be used in any area where the environment 
does not exceed the specifications of Table B-2.

Avoid exposing the 620A 4R to extremes of 
temperature which will affect accuracy and 
shorten battery life-span. 

Содержание 620A 4R

Страница 1: ...AUSTIN TEXAS 78737 USA CAGE CODE 1CRL2 PHONE 512 858 4045 or 1 800 350 5105 from the USA e mail info amptec com FAX 512 858 4340 or 1 800 430 5440 from the USA website http www amptec com 620A 4R Batt...

Страница 2: ...ARRANTY ________________________________________________________ Permission and a return authorization RMA number must be obtained directly from AMPTEC s customer service department via phone FAX or e...

Страница 3: ...Charger Power Requirements A 5 Setup and Use CHAPTER B 620A 4R FAILSAFE OHMMETER SPECIFICATIONS Table B 2 Specifications CHAPTER C REPLACEMENT OPTIONAL ANDACCESSORYITEMS C 1 Available Accessories and...

Страница 4: ...ER F ROUTINE MAINTENANCE F 1 General F 2 Required Test Equipment F 3 Calibration Procedure F 4 Battery Replacement Instructions CHAPTER G COMPONENT PARTS LISTS 620 400 Parts Assembly list CHAPTER H SC...

Страница 5: ...eable batteries reach a low charge level a negative sign will appear on the display The 620A 4R has a battery monitoring circuit that indicates it is time to plug in the battery charger A 2 Receiving...

Страница 6: ...d is open The 620A 4R has a way to remove it s lid completely for example when used in an indoor laboratory environment With the 620A 4R open remove the units two hinge pins hex type by twisting them...

Страница 7: ...entioned earlier of course the AC DC Battery Charger must be plugged into the 620A 4R s notched connector to facilitate charging the batteries Although the batteries are fully charged prior to shipmen...

Страница 8: ...d 35 C 50 C Instrument Display 20 000 count 4 digit Super Bright Light Emitting Diodes LED Over Range Indication select next higher range 620A 4R Display flashes Measurement Update Rate Approximately...

Страница 9: ...er that verifies each individual replacement battery voltage charge level inspector signature along with the inspection date CHAPTER C REPLACEMENT OPTIONALAND ACCESSORY ITEMS 3 2 Test Lead Sets and Pr...

Страница 10: ...Terminal Kelvin Mini probes photo shown above Each probe has two spring loaded gold plated steel tips with 0 18 separation one red and one black handheld probe Excellent general purpose 4 Wire Kelvin...

Страница 11: ...anana leads which provide a 4 wire Kelvin connection from the adapter to the 620A Ohmmeter No lead resistance errors In some applications i e AIM 7 Missile harnesses regular alligator clip leads canno...

Страница 12: ...s just below the display on the left see item 11 of the 620A 4R Front Panel Diagram When a given range is selected an indicator LED informs the user Also note that a resistance range must be selected...

Страница 13: ...rews have been removed for calibration access 4 Test Lead Connection Panel Jacks The 48 long dual banana Alligator Test Leads plug directly into the Voltage High and Voltage low and Current high and C...

Страница 14: ...item 7 of the 620A 4R Diagram The fuse protects internal batteries from excessive charging currents Replace blown fuses with the same type and rating only The Charging System The slimline AC to DC bat...

Страница 15: ...shielded In addition custom probes and test lead sets are optionally available for customized wiring connections kelvin clips cables terminated with spade lugs and special banana jacks All AMPTEC ohm...

Страница 16: ...512 447 7456 International Overseas or FAX 512 447 7455 email service amptec com D 7 Failsafe Operation The AMPTEC 620 Series of Igniter Testers or failsafe ohmmeters incorporate a constant current s...

Страница 17: ...50 VDC To make this adjustment first remove the fuse from the fuseholder With an adjustable DC power supply set the power supply output to be 4 50 VDC Be sure to observe power supply polarity Connect...

Страница 18: ...cate the batteries need charging or fuse needs replacing If the 620A 4R exhibits problems that cannot be eliminated by reviewing Chapters B and D the following guidelines have been established to help...

Страница 19: ...or move the replacement part until the solder has cooled Cold solder and bad solder joints can cause more problems Use the chassis ground connect to the common terminal of the functional test section...

Страница 20: ...the analog part and the ICL71C03 takes care of the digital part of the 4 digit 20 000 count dual slope conversion COUNTERS DISPLAY COMPARATOR BUFFER MULTIPLEXER ZERO CROSSING DETECTOR INTEG UUT CONTRO...

Страница 21: ...EGRATOR INTEGRATOR COMPARATOR COMPARATOR CAZ CAZ CSTRAY CSTRAY RINT RINT FIGURE E 2B PHASE ll INTEGRATE INPUT FIGURE E 2C PHASE lll AND DE INTEGRATE CINT CINT VIN VIN CREF CREF 3 3 1 1 2 2 6 6 5 5 4 4...

Страница 22: ...shed in Phase I The time or number of counts required to do this is proportional to the input voltage E 4 1 Reference Voltage The precision reference voltage required to do the A D conversion is devel...

Страница 23: ...t as it was when the output terminals were shorted The current through R is 10 L milliamperes just as it was through the jumper when the output terminals were shorted E 6 Failsafe Design Reference to...

Страница 24: ...wer Supply Scheme The 5 volt power supply is provided directly by the batteries for driving the LED displays and digital logic The 5VD is used for driving IC8 the low battery detection circuit The 15V...

Страница 25: ...range push buttons covered by silicone rubber boots and their corresponding panel mount LEDS IC 2 is a latch that will latch its output to the corresponding range push button input Its output turns o...

Страница 26: ...AMPTEC 620 package may have too much resistance offset for the actual calibration of 620 Igniter Testers versions fitted with a low 2 0 Ohm range Contact AMPTEC for information on the 305 48 long 4 w...

Страница 27: ...n Return the calibration access screws to the perimeter of the 620 main top panel 2 Adjust RV1 for a display indication of 100 00 3 Check the 2 0 KOhm range with the 1000 Ohm Standard Resistor Check t...

Страница 28: ......

Страница 29: ...3 D104 and D202 D201 D203 Relay Board 04 30015 1 620A Relay Board DWG 620A 702 REVB RL1 05 03006 1 Relay Aromat S2EB 5V CHG1 05 79393 1 Slimline 9 VDC Battery Charger AMPTEC Modified Volgen Model LV 1...

Страница 30: ...rostat 50KW DK RV6N503C ND Allied 753 2503 HNDL1 05 77201 2 Handles Front Panel Mount Type Allied 839 9119 STF1 05 77251 2 Standoff Box of 100 Allied 839 2357 KB1 05 77311 1 Knob for shaft of the zero...

Страница 31: ...ly Rev D 620 601 2 1 page 620 Display Board Assembly 620 070 3 pages 620 Ohmmeter Schematics 1 DVM1 Rev D 2 I_SRC Current Source Rev D 3 DISP Display Rev D 620 077 1 page 620 Relay PCB Schematic Rev A...

Страница 32: ...bon Cable Header routes to the Top Panel D6 Q6 D7 Q7 D8 8 Q8 470W R6 470W R10 620 RELAY BOARD Drawing 620 077 Cage Code 1CRL2 RESEARCH AMPTEC Page 1 of 1 INPUTS OUTPUT OE LE D Q L H H H L H L L H X X...

Страница 33: ...620 FRONT PANEL RANGE SWITCH SCHEMATIC Drawing 620 078 RESEARCH AMPTEC Page 1 of 1 Front Panel LED Indicators 2KW Range Indicator 200W Range Indicator Revision A Scale none...

Страница 34: ...D1 C1 C2 D2 RV1 D3 R4 R9 R6 R14 RV201 R13 R3 R2 R1 R15 Q5 R7 C3 C5 C4 R8 IC3 C11 RV2 C12 IC4 RN1 Q4 Q3 Q2 Q1 D5 R17B C8 IC9 C10 S1 D4 R225 R224 R223B C13 RV3 P1 R20 R19 IC8 C9 R16 R18 IC10 IC11 R23 R2...

Отзывы: