ROUTINE MAINTENANCE
45
OTOWAVE 302+ INSTRUCTION FOR USE
6.
ROUTINE MAINTENANCE
6.1.
GENERAL MAINTENANCE PROCEDURES
The performance and safety of the instrument will be maintained if the following recommendations for care and
maintenance are observed:
1.
It is recommended that the instrument go through at least one annual service, to ensure that the acoustical,
electrical and mechanical properties are correct. This should be carried out by an authorised service centre in
order to guarantee proper service and repair.
2.
Observe that no damage is present to the insulation of the mains cable or the connectors and that it is not
exposed to any kind of mechanical load that could involve damage.
3.
To ensure that the reliability of the instrument is maintained, we recommend that the operator at short
intervals, for instance once a day, performs a test on a person with known data. This person could be the
operator.
4.
If the surface of the instrument or parts of it is contaminated, it can be cleaned using a soft cloth moistened
with a mild solution of water and detergent or similar. Always disconnect the mains power adaptor during the
cleaning process and be careful that no fluid enters the inside of the instrument or accessories.
5.
After each patient examination, ensure that there has been no contamination to the parts touching the patient.
General precautions must be observed in order to avoid cross-contamination of disease from one patient to
another. Water should be used for frequent cleaning, but in the case of severe contamination it may be
necessary to use a disinfectant.
•
Before cleaning always switch off and disconnect from the power supply
•
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
•
Do not allow liquid to come into contact with the metal parts inside the earphones/headphones
•
Do not autoclave, sterilise or immerse the instrument or accessory in any fluid
•
Do not use hard or pointed objects to clean any part of the instrument or accessories
•
Do not let parts that have been in contact with fluids dry before cleaning
•
Rubber ear-tips or foam ear-tips are single use accessories
Recommended cleaning and disinfection solutions:
•
Warm water with mild, nonabrasive cleaning solution (soap)
•
70% isopropyl alcohol only on hard cover surfaces
CAUTION