USING THE OTOWAVE 302+
25
OTOWAVE 302+ INSTRUCTION FOR USE
RESET ONLY 1 SETTING (SWEEP OR REFLEX)
1.)
To reset just one setting, access the sweep or reflex settings. In there you will find a
DEFAULT
function.
2.)
Choose the profile you want to reset: the current selected one or all available profiles. After selection, confirm
the overwrite.
4.4.5.
DISPLAY MODE
To view the current selected profile, ensure that the
MAIN MENU
is displayed and
then press and hold the function button on the probe to display the
TEST
SETTINGS
screen as shown below.
1
Profile:
Current selected Profile
2
Probe:
Frequency being used for tympanometry (226 or 1000 Hz).
3
Reflexes:
Reflexes tested (ipsilateral or contralateral or both).
4
Frequencies and
Level:
Selected frequencies for reflex testing (marked with a
when selected), max.
level to test and step size.
5
Probe #:
Serial number of probe used.
6
Contra #:
Serial number of contra phone calibrated to equipment.
2
3
4
1
5
6