UNPACKING AND INSTALLATION
16
OTOWAVE 302+ INSTRUCTION FOR USE
2.7.
LIGHT INDICATORS
The indicators on the Otowave 302+ and the probe show the status of the system.
STATUS
LED B
LED C
PROBE
Otowave 302+ turned off
Off
Off
Off
Idle, test completed or test cancelled
On
Off
Green on
Insert probe or remove probe (refer to display for details)
Flashing
(fast)
Flashing
(fast)
Colour alternating
(Green / Yellow)
Ensure probe is held steady while an ear seal is obtained
Off
Flashing
(slow)
Yellow flashing
(slow)
Testing - tympanogram and/or reflex measurement
Flashing
(slow)
Off
Green flashing
(slow)
2.8.
CONTRALATERAL TRANSDUCER
1 Ear tip
Ear tip to be placed on probe tip of contra phone
2 Probe tip
Probe tip screwed onto contra phone
3 Plug
Connector to
CONTRA
socket on Otowave 302+
The contralateral transducer is used when it is required to provide a reflex stimulus to the opposite ear to that being
tested with the main probe assembly. For use it should be connected to the
CONTRA
socket on the base unit and fitted
with a new ear tip.
The contralateral probe tip may be replaced if necessary (e.g. if damaged). To remove the contralateral probe tip,
carefully unscrew it from the body of the transducer. Carefully fit the replacement part and make sure that it is screwed
home firmly but do not over-tighten. Do not use any tools to tighten the contralateral probe tip.
2.9.
CHOOSING THE CORRECT EAR TIP SIZE
Amplivox YouTube video is available for assistance on how to choose the right ear tip.
The probe tip must be fitted with a new ear tip before it is presented to a
patient’s ear canal. If a contralateral reflex
stimulus is to be applied, fit a new ear tip to the contralateral transducer before presenting it to the patient’s opposite
ear canal.