background image

User Guide - ATU 

 

 

5 von 8

 

Technical Data are subject to change without 

notice

 

User Guide ATU v.5.0 Test 

kit.docx-5.0

 

 

 

      

 

 
Otherwise you will see a communication error. 

 

 

5  Position measurement (ATU-1 and ATU-2) 

 

Check the geometrical mounting conditions (use Mounting Instructions 

https://www.amo-

gmbh.com/

 

If the ATU Box was connected successfully, you can start getting position with click at “Start” 
button. 

 

If something  goes  wrong  during  measurements,  you  will see  error messages. Please  follow the 
instructions and try again. 
 

To stop measurement process click “Stop” button. 

If  you  need  to  set “0” position,  click “Set  current  position  as “0”” button. The  button  will change 
name to “Reset “0” position” and the value in measurement data field will be set as 0. The value 
of “0” position will be saved until you reset it by clicking the same button or close the programm. 

Before closing programm, please break connection with ATU Box by clicking “Device connected” 
button. 

 
 

If  an  error  occurs  the  textbox  will  change  colors  to  red  until  the  error  is  gone.  The  “ERROR 
Acknowledge” button stays red until you press the button while there are no active errors. 

Содержание ATU

Страница 1: ...v 5 0 Test kit docx 5 0 AMO GmbH ATU Test kit Contents 1 Generalities 2 2 Software Installation 3 3 Hardware 4 4 Software Manual 4 5 Position measurement ATU 1 and ATU 2 5 6 Measuring of signal accura...

Страница 2: ...ATU 1 connection scheme Pic 2 ATU 2 connection scheme Minimal PC requirements Processor 1GHz RAM 512 MB Operating system Windows XP or higher Hardware ATU Box ATU 1 or ATU 2 USB cable ATU to PC Conne...

Страница 3: ...to change without notice User Guide ATU v 5 0 Test kit docx 5 0 ATU 1 Test Kit ATU 2 Test Kit 2 Software Installation Download software from official site of AMO GmbH http www amo gmbh com Launch the...

Страница 4: ...ized and installed the ATU Box Connect ATU Box with measuring head using connection scheme Page 2 4 Software Manual Start the Software with desktop icon or file AMO_ATU exe in installed folder Select...

Страница 5: ...sages Please follow the instructions and try again To stop measurement process click Stop button If you need to set 0 position click Set current position as 0 button The button will change name to Res...

Страница 6: ...f periods Here you have an opportunity to set measurement speed precision between 1 and 30 and minimal periods quantity for measurement between 50 and 2100 The precision of evaluation is higher for un...

Страница 7: ...www amo gmbh com If the ATU Box was connected successfully you can start getting functional reserve value with click at Start button The Functional Reserve value is updated during the movement After...

Страница 8: ...BiSS C Encoder registers Get Register button reads all the registers form the current slave By double clicking on a value the user is able to type a new value for the register All values are in hexade...

Отзывы: