AML Oceanographic Keller PA-10 Скачать руководство пользователя страница 13

User Manual for AML Oceanographic’s Micro Pressure Sensor 

 

 

AML OCEANOGRAPHIC 

In-situ Real-time Detection Measurement

 

13 

4 MAINTENANCE 

 

 

The Micro Pressure Sensor has been designed to minimize user maintenance. To keep the 

instrument in top condition the following maintenance is required: 

 

 

After each deployment, the Micro Pressure Sensor case should be washed thoroughly with 

fresh water. 

 

Rinse (with freshwater) and dry the connector if contamination with saltwater is suspected.  

Saltwater  inside  the  connector  will  eventually  corrode  the  contacts  causing  intermittent 

communications. 

 

Caution:

 

 

The pressure sensor is fragile. Contact with a sharp object to the pressure sensor face 

could damage it. 

 

 

4.2 Main O-Ring Maintenance 

 

Normal use of the Micro Pressure Sensor will not require the main O-ring to be changed. 

However, the main O-ring may be inspected periodically and greased to keep the O-ring 

in good condition.  If during an inspection damage is suspected please do the following. 

 

Note: Do not use sharp objects to remove the main O-ring from its groove. 

  

 

1.

 

Using the method outlined in 

Figure 4.0

, use the thumb and index finger to remove 

the main O-ring from its groove. 

2.

 

Clean the O-ring using a tissue or clean cloth. 

3.

 

Inspect the O-ring for cuts or nicks. An eyepiece or magnifying glass will help. 

Replace if necessary. 

4.

 

Clean the O-ring groove located on the end cap with a swab or tissue. 

5.

 

Inspect the O-ring groove for scratches or dirt. 

6.

 

Lightly coat the O-ring with silicon grease and replace it onto the end cap. 

7.

 

Clean the O-ring surface located on the inside of the pressure housing with a swab or 

tissue. 

8.

 

Inspect the O-ring surface for dirt or scratches. 

9.

 

Lightly coat the O-ring surface on the inside of the pressure housing with silicone 

grease. 

 

 

 

 

Содержание Keller PA-10

Страница 1: ...y British Columbia CANADA V8L 5X6 Tel 1 250 656 0771 Toll Free 1 800 663 8721 Fax 1 250 655 3655 Email sales AMLoceanographic com www AMLoceanographic com Micro Pressure Sensor ____________________ Us...

Страница 2: ...NECTING THE MICRO PRESSURE SENSOR TO A COMPUTER 6 2 3 POWERING THE MICRO PRESSURE SENSOR 6 3 COMMUNICATING WITH THE MICRO PRESSURE SENSOR 7 3 1 CONFIGURING HYPERTERMINAL FOR COMMUNICATIONS WITH THE MI...

Страница 3: ...nts The warranty does not apply if the instrument has been damaged by accident or misuse and is void if repairs or modifications are made by any other than authorized personnel This warranty is the on...

Страница 4: ...autobaud communications Baud rate 2400 to 38400 8 data bits no parity 1 stop bit Data type ASCII text Sample Rate 25 scans per second Power Isolated External DC supply Voltage range 8 26 Volts DC Curr...

Страница 5: ...uter The instrument s output is standard ASCII RS 232 The baud rate is automatically determined when the sensor receives an ENTER or RET The data output may be configured to display either unprocessed...

Страница 6: ...on device An AML supplied data cable will link the two Refer to the wiring diagram in Appendix C At one end of the cable is a DB25 or DB9 female connector that plugs into the computer s serial port an...

Страница 7: ...properly and the appropriate baud rate serial port combination chosen the instrument will return the prompt Once the prompt appears the Micro Pressure Sensor is ready to accept instructions Refer to...

Страница 8: ...AML Oceanographic s Micro Pressure Sensor AML OCEANOGRAPHIC In situ Real time Detection Measurement 8 Figure 3 2 7 Choose the Baud rate 19200 8 Data bits No Parity 1 Stop bit and no Flow control Figur...

Страница 9: ...e Sensor can display data in two modes real or raw Real mode displays data that has been calculated to produce engineering units 3 2 1 Raw Data Format Data displayed in raw mode is strictly a numeric...

Страница 10: ...ill set the instrument to output RAW uncorrected data when using the MONITOR SCAN or DUMP commands Command REAL RE This command will set the instrument to output Real corrected engineering data when u...

Страница 11: ...mand SET DETECT ab SE D ab Where a a Hex number between 0 F b a Hex number between 0 F This command sets the detection mode of the instrument on power up The DETECT command can be used to set the Micr...

Страница 12: ...e serial number of the instrument This command is intended to facilitate integration into OEM equipment To display the instruments serial number use the VER command Example set sn 7245 To display the...

Страница 13: ...main O ring to be changed However the main O ring may be inspected periodically and greased to keep the O ring in good condition If during an inspection damage is suspected please do the following No...

Страница 14: ...surement 14 O Ring To remove an O ring squeeze the end cap between thumb and finger and slide forward the O ring will bulge and can be easily removed with the other hand Do not be tempted to dig the O...

Страница 15: ...ence points at different temperatures This data is applied to a curve fitting algorithm to produce calibration coefficients for both pressure and temperature effects Each set of coefficients is perman...

Страница 16: ...nicate with the computer Solutions Check the power supply for correct voltage setting 8 26 Volts DC and or current limiting The serial port chosen is incorrect Most IBM PC compatible computers have on...

Страница 17: ...User Manual for AML Oceanographic s Micro Pressure Sensor AML OCEANOGRAPHIC In situ Real time Detection Measurement 17 APPENDIX B General Layout...

Страница 18: ...NG DIAGRAM MICRO8 RS232 MICRO SENSOR PACK 9 Pin D Connector 25 Pin D Connector 5 1 N C Screen Blue 6 7 5 1 Power supplied by customer supplied 12VDC Power or 12VDC AML Battery Pack optional PN SUB G01...

Отзывы: