AML Oceanographic Keller PA-10 Скачать руководство пользователя страница 12

User Manual for AML Oceanographic’s Micro Pressure Sensor 

 

 

AML OCEANOGRAPHIC 

In-situ Real-time Detection Measurement

 

12 

Example 1: 

 

>set detect 97  
 

 In the above example, the Micro CTD will try 9 times to auto baud. If the instrument is not 
able to establish a baud rate, it will default to 38400 baud. 
 

Example 2: 

 

>set detect 05 

 
In the above example, the Micro CTD will 

not 

try to auto baud, but will simply default to 

9600 baud on power up. 

 

Command: 

SET SN

 nnnn 

 

This  command  sets  the  serial  number  of  the  instrument.  This  command  is  intended  to 
facilitate integration into OEM equipment.  To display the instruments serial number use the 
VER command. 

 

Example:  

 

 

>set sn 7245 

 

 

To display the serial number, 

>ver 
Micro CTD Sensor V2.07 SN:7245-CTD 
Copyright© 2001-2003, AML Oceanographic 
0MB of Memory 

 

Warning:  Changing  the  instruments  serial  number  will  adversely  effect  the    

operation of Smart Talk or ISS software. 

 

 

Command: 

DIS C 

(Display) 

 

 

This command will display the instrument’s current calibration coefficients.  

 

 

 

 

>Dis c   

returns the instruments calibration coefficients 

 

Command: 

DIS S 

 

This command displays the current scan rate.  

Example: 

 

>DIS S  

 

 

sample rate is 1 seconds 

 

 

 

 

 

Содержание Keller PA-10

Страница 1: ...y British Columbia CANADA V8L 5X6 Tel 1 250 656 0771 Toll Free 1 800 663 8721 Fax 1 250 655 3655 Email sales AMLoceanographic com www AMLoceanographic com Micro Pressure Sensor ____________________ Us...

Страница 2: ...NECTING THE MICRO PRESSURE SENSOR TO A COMPUTER 6 2 3 POWERING THE MICRO PRESSURE SENSOR 6 3 COMMUNICATING WITH THE MICRO PRESSURE SENSOR 7 3 1 CONFIGURING HYPERTERMINAL FOR COMMUNICATIONS WITH THE MI...

Страница 3: ...nts The warranty does not apply if the instrument has been damaged by accident or misuse and is void if repairs or modifications are made by any other than authorized personnel This warranty is the on...

Страница 4: ...autobaud communications Baud rate 2400 to 38400 8 data bits no parity 1 stop bit Data type ASCII text Sample Rate 25 scans per second Power Isolated External DC supply Voltage range 8 26 Volts DC Curr...

Страница 5: ...uter The instrument s output is standard ASCII RS 232 The baud rate is automatically determined when the sensor receives an ENTER or RET The data output may be configured to display either unprocessed...

Страница 6: ...on device An AML supplied data cable will link the two Refer to the wiring diagram in Appendix C At one end of the cable is a DB25 or DB9 female connector that plugs into the computer s serial port an...

Страница 7: ...properly and the appropriate baud rate serial port combination chosen the instrument will return the prompt Once the prompt appears the Micro Pressure Sensor is ready to accept instructions Refer to...

Страница 8: ...AML Oceanographic s Micro Pressure Sensor AML OCEANOGRAPHIC In situ Real time Detection Measurement 8 Figure 3 2 7 Choose the Baud rate 19200 8 Data bits No Parity 1 Stop bit and no Flow control Figur...

Страница 9: ...e Sensor can display data in two modes real or raw Real mode displays data that has been calculated to produce engineering units 3 2 1 Raw Data Format Data displayed in raw mode is strictly a numeric...

Страница 10: ...ill set the instrument to output RAW uncorrected data when using the MONITOR SCAN or DUMP commands Command REAL RE This command will set the instrument to output Real corrected engineering data when u...

Страница 11: ...mand SET DETECT ab SE D ab Where a a Hex number between 0 F b a Hex number between 0 F This command sets the detection mode of the instrument on power up The DETECT command can be used to set the Micr...

Страница 12: ...e serial number of the instrument This command is intended to facilitate integration into OEM equipment To display the instruments serial number use the VER command Example set sn 7245 To display the...

Страница 13: ...main O ring to be changed However the main O ring may be inspected periodically and greased to keep the O ring in good condition If during an inspection damage is suspected please do the following No...

Страница 14: ...surement 14 O Ring To remove an O ring squeeze the end cap between thumb and finger and slide forward the O ring will bulge and can be easily removed with the other hand Do not be tempted to dig the O...

Страница 15: ...ence points at different temperatures This data is applied to a curve fitting algorithm to produce calibration coefficients for both pressure and temperature effects Each set of coefficients is perman...

Страница 16: ...nicate with the computer Solutions Check the power supply for correct voltage setting 8 26 Volts DC and or current limiting The serial port chosen is incorrect Most IBM PC compatible computers have on...

Страница 17: ...User Manual for AML Oceanographic s Micro Pressure Sensor AML OCEANOGRAPHIC In situ Real time Detection Measurement 17 APPENDIX B General Layout...

Страница 18: ...NG DIAGRAM MICRO8 RS232 MICRO SENSOR PACK 9 Pin D Connector 25 Pin D Connector 5 1 N C Screen Blue 6 7 5 1 Power supplied by customer supplied 12VDC Power or 12VDC AML Battery Pack optional PN SUB G01...

Отзывы: