Installation
6. Connect the hanging bracket to the headwall support (
Figure 8
) using the provided self-drilling screws
(
Hardware 2
). Use the number of screws listed below based on section size:
a. 19" to 32" wide section = 2 screws
b. 32.1" to 48" wide section = 4 screws
7. Connect the medical gases to the FlexWall Engine.
NOTE:
The ends of Hose Assemblies have gas-specific adapters meant to plug into the FlexWall Engine.
WARNING:
Orient and secure gas hoses so that they do NOT create a “valley” (i.e., a “U”-shape) (
Figure 9
).
NOTE:
Medical gas hoses are terminated as per shop drawings.
Figure 8
Ceiling Line
Floor Line
Stud
Bottom Track
Top Track
Blocking
Screw provided by Amico
(Hardware 2)
Brackets secured to the
wall (hardware by others)