background image

     www.amico.com     

21

Environmental Conditions

Operation

Minimum

Maximum

Temperature

+10°C

+40°C

Relative atmospheric humidity

30%

75%

Air pressure

700 hPa

1060 hPa

Transport/Storage

Minimum

Maximum

Temperature

-10°C

+50°C

Relative atmospheric humidity

20%

90%

Air pressure

700 hPa

1060 hPa

NOTE: 

It is recommended that the lift be brought back to an area kept at room temperature between uses. For 

example after transport in a vehicle do not leave the lift in the trunk or backseat.

Also, try to keep the lift from staying in rooms with a metallic roof where temperatures may increase the allowable 
operating temperature.

References on the Package

RH

20% - 90%

P

700hPa - 1060hPa

-10°C

(14°F)

+50°C

(122°F)

Temperature range

for transport

and storage

Atmospheric

humidity

for transport

and storage

Air pressure

for transport

and storage

Содержание GoLift Portable450

Страница 1: ...Installation and Operation Instructions Portable450 Portable450 ...

Страница 2: ...e Amico Track 10 Labeling the Tracks 11 Using the Reacher Bar 12 Magnetic Connection of the Hand Control to the GoLift Portable450 12 Operating the GoLift Portable450 13 Charging the GoLift Portable450 13 LED and Buzzer Functions 14 Emergency Stop 15 Emergency Lowering 15 Manual Emergency Lowering 15 Overspeed Cam 15 Cleaning and Disinfection 16 Troubleshooting 16 Inspection and Maintenance 17 GoL...

Страница 3: ...ook under the cover of this compact lift you will find an impressive set of all metal gears and state of the art battery technology that will allow the caregiver to safely and effortlessly transfer a patient weighing up to 450 lbs on a single lift The GoLift Portable450 weight capacity is 450 lbs This weight limit must not be exceeded The Amico GoLift Portable450 is a fixed lift and can be install...

Страница 4: ... Corporation Email amo service amico com 1 877 462 6426 M A D E IN NORTH AM E R I C A Charger Wall Industries 36V 1 11A 40W Duty Cycle 1 Min ON 9 Min OFF Serial Mfg Date GLP450 9999 450 lbs 205 kg IP23 7 315 A B 3 3 2 2 1 1 Back Front 0 8341 1 6256 WARNING This symbol is intended to alert the user of hazard or unsafe practices which could result in serious bodily harm MARKINGS The GoLift Portable4...

Страница 5: ...ould be checked to ensure that they are in good working order Check for signs of tear or fraying prior to use Report any unusual wear or damage immediately Amico Mobility will not be responsible for any damage caused by misuse neglect or purposeful destruction of the lift and its associated components Do not attempt to modify alter the Amico GoLift Portable450 Do not in any circumstance exceed the...

Страница 6: ...g Lasting NiMH 1 x 24V 1 6Ah Battery Charger Input 100 240VAC Output 36VDC 40 Watt 1 0A Hand Control IPX4 Hand Control protects from splashing water no matter the direction Safety Emergency Stop Emergency Lowering Device Independent Circuit Manual Lowering Upper Limit Detection Lower Limit Detection Slack Tape Sensor Free Fall Brake Over Speed Governor Low Battery and Dead Battery Alarms Soft Star...

Страница 7: ...able450 Emergency lowering Durable all metal gears Unit controls Quick magnetic hand control connection Wired hand control Lifts up to 450 lbs GoLift Portable450 weighs just 6 lbs Back of the GoLift Portable450 Emergency lowering Emergency stop ...

Страница 8: ...ble450 Start with the uppermost set of straps from the back and then take the lowest set of straps from the legs Be careful when attaching the lifting sling on the hooks Check that the straps have been completely through the opening and into place in the Portable lift hooks When pressing the up button to lift the patient check again that all straps remain correctly placed Lifting to and from a sea...

Страница 9: ...he bed so that the patient is sitting up Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of slings should use the Amico GoLift Portable450 Important Plan the move and avoid leaving the patient in the sling unattended Before lifting check that the patient is completely free of their surroundings The patient s head arms hands and feet must ...

Страница 10: ...m and place the endcap back on the arm 5 Fasten all endstops tightly using an adjustable wrench NOTE If you are installing the Amico GoLift Portable450 in an existing track system you must ensure that the max load of the track system is equal or higher than the max load of the Amico GoLift Portable450 6 To disconnect the lift from the trolley open up the hook in the carabiner to release Installing...

Страница 11: ...ere the label is to be attached using one of the recommended cleaning agents see Cleaning and Disinfection section Allow the track to dry before peeling the label from the backing and attaching to either side of the track With a permanent writing material fill in the date that the track was certified using a permanent marker Example Track Label Labeling the Tracks 1 877 462 6426 info amico com Ami...

Страница 12: ...lish this step otherwise it will not be possible to hook on 4 Slide the top hook of the Reacher Bar through the trolley Make sure there is enough slack on the lift strap to accomplish this step otherwise it will not be possible to hook on 5 The lift has now been securely hooked onto the trolley The Reacher Bar remains in position until the lift is removed Using the Reacher Bar Open carabiner to at...

Страница 13: ...ies The GoLift Portable450 may remain connected to the charger indefinitely since the charger has a built in regulator eliminating the danger of overcharging As a general rule it is recommended that the lift be raised to a height that will not interfere with anything or anyone while the lift is not in use Do not drain the batteries excessively This will dramatically reduce the lifespan of the batt...

Страница 14: ...LED Main Lift Buzzer Emergency down Solid Red 2 beeps per second Maintenance Alarm At 1001 lifts the LED will flash Green indicating the unit is due for a maintenance check To reset press Up button then Down button then Up button then Down button and then hold both buttons for 10 seconds If successful the LED turns Red Green and Orange The buzzer beeps three times Please note that a reset can be d...

Страница 15: ... failure situations the patient may be lowered by pressing and holding down the RED emergency lowering button located on the top of the lift Continue to press down until the patient is safely lowered to the desired position The lift will beep as you continue to press down on the red button and will continue beeping until the button is released after the desired lowering has been achieved NOTE The ...

Страница 16: ... 5 Virox Accel Tb Clean the inside of the tracks every 4 months using a damp cloth to ensure a smooth rolling surface for the trolley Troubleshooting Should problems arise with the use of the Amico GoLift Portable450 review the following chart Find the fault and complete the recommended solution If the fault is not found and or the solution does not correct the problem contact Amico Mobility Fault...

Страница 17: ...e gears for any broken or worn teeth Verify the overspeed cam is operating freely Check emergency stop button Check emergency lowering device Annual load test with SWL maximum safe working load Ensure the endstops are installed Reapply grease to the gears In accordance to the ISO 10535 Standard Hoists for the transfer of disabled persons Requirements and test methods an inspection should be perfor...

Страница 18: ...SLG GO BASIC MSH L SLG GO BASIC MSH XL GoLong Polyester SLG GO LONG WV 42X78 SLG GO LONG WV 56X78 GoLong Mesh SLG GO LONG MSH 42X78 SLG GO LONG MSH 56X78 GoHygiene Polyester SLG GO HYGIENE WV S SLG GO HYGIENE WV M SLG GO HYGIENE WV L SLG GO HYGIENE WV XL GoComfort Polyester SLG GO COMFORT WV S SLG GO COMFORT WV M SLG GO COMFORT WV L SLG GO COMFORT WV XL Guidelines For Choosing Your Sling Size S M ...

Страница 19: ...______________________ DESCRIPTION BY DATE 5 4 3 1 55 East Wilmot St Richmond Hill Ontario L4B 1A4 CANADA Tel 905 764 0800 Fax 905 764 0862 DESCRIPTION PART NO SHEET OF DRAWING NO 1 1 PROJECTION PRIVATE AND CONFIDENTIAL THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN CONFIDENCE SUBJECT TO RETURN UPON REQUEST AND WITH THE UNDERSTANDING THAT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF ...

Страница 20: ...x 2 63 x 2 36 6100 mm x 67 mm x 60 mm 51 lbs 23 1 kg Double TRK DBL 95 95 x 2 63 x 3 55 2413 mm x 67 mm x 90 mm 25 lbs 11 4 kg TRK DBL 120 120 x 2 63 x 3 55 3050 mm x 67 mm x 90 mm 32 lbs 14 5 kg TRK DBL 150 150 x 2 63 x 3 55 3810 mm x 67 mm x 90 mm 40 lbs 18 1 kg TRK DBL 180 180 x 2 63 x 3 55 4570 mm x 67 mm x 90 mm 48 lbs 21 8 kg Triple TRK TRP 120 120 x 2 63 x 5 75 3050 mm x 67 mm x 146 mm 38 l...

Страница 21: ...at the lift be brought back to an area kept at room temperature between uses For example after transport in a vehicle do not leave the lift in the trunk or backseat Also try to keep the lift from staying in rooms with a metallic roof where temperatures may increase the allowable operating temperature References on the Package RH 20 90 P 700hPa 1060hPa 10 C 14 F 50 C 122 F Temperature range for tra...

Страница 22: ...ishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operat...

Страница 23: ...1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT...

Страница 24: ...in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagati...

Страница 25: ...0 23 0 1 0 38w 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz an...

Страница 26: ...tions Corporation 85 Fulton Way Richmond Hill ON L4B 2N4 Canada 600 Prime Place Hauppauge NY 11788 USA Tel 905 764 0800 Fax 905 764 0862 Warranty Policy Mobility Solutions The primary purpose of our ceiling lift system is to safely lift transfer and reposition a patient with as little effort as possible for the caregiver regardless of the room type The Amico Mobility Solutions Corporation s GoLift...

Страница 27: ...www amico com 27 Notes ...

Страница 28: ... B East Beaver Creek Road Richmond Hill ON L4B 1G6 Canada 600 Prime Place Hauppauge NY 11788 USA Toll Free Phone 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email amo service amico com www amico com AJW Technology Consulting GmbH Königsallee 106 40215 Dusseldorf Germany ...

Отзывы: