Electro-Magnetic Compliance Data For Amico GoLift Portable

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below. 
The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.

Emissions test 

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment – Guidance

√Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to 80 MHz

3 V/m
80 MHz to 2,5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications 
equipment should be used no closer to any 
part of the [ME EQUIPMENT or ME SYSTEM], 
including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of the 
transmitter.

Recommended separation distance 

d = 1.2 √P
d = 1.2 √P  80 MHz to 800 MHz
d = 2.3 √P  800 MHz to 2,5 GHz

where 

P

 is the maximum output power rating 

of the transmitter in watts (W) according to 
the transmitter manufacturer and 

d

 is the 

recommended separation distance in meters (m). 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, 

a. 

should be less than the compliance level in 

each frequency range. 

b.

Interference may occur in the vicinity of 
equipment marked with the following symbol:

 

NOTE 1:

 At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2:

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people. 

a.

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. 
If the measured field strength in the location in which the [ME EQUIPMENT or ME SYSTEM] is used exceeds the applicable 
RF compliance level above, the [ME EQUIPMENT or ME SYSTEM] should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as re-orienting or relocating the [ME EQUIPMENT or  
ME SYSTEM].

b. 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Содержание Go Lift Portable 450

Страница 1: ...Installation and Operation Instructions w w w a m i c o c o m Portable450 Portable450...

Страница 2: ...on the Amico Track 10 Using the Reacher Bar 11 Connecting the Hand Control to the GoLift Portable 11 Operating the GoLift Portable 12 Charging the GoLift Portable 12 LED and Buzzer Functions 13 Emerge...

Страница 3: ...look under the cover of this compact lift you will find an impressive set of all metal gears and state of the art battery technology that will allow the caregiver to safely and effortlessly transfer a...

Страница 4: ...ndards Standard s CAN CSA C22 2 No 60601 1 08 Medical Electrical Equipment Part 1 General requirements for basic safety and essential performance ISO10535 2006 Hoists for the transfer of disables pers...

Страница 5: ...checked to ensure that they are in good working order Check for signs of wear or fraying prior to use Report any unusual wear or damage immediately Amico Mobility will not be responsible for any damag...

Страница 6: ...ntrol 1m water resistance IP23 Lift protected from fingers and water spray less 60 degrees from vertical Batteries High Capacity Long Lasting NiMH 1 x 24V 1 6Ah Battery Charger 100 240 VAC 40 Watt Han...

Страница 7: ...Anatomy of the GoLift Portable Emergency lowering stop feature Unit controls Quick magnetic hand control connection Wireless hand control Durable all metal gears Lifts up to 450 lbs GoLift Portable we...

Страница 8: ...hooks of the GoLift Portable Start with the uppermost set of straps from the back and then take the lowest set of straps from the legs Be careful when attaching the lifting sling on the hooks Check t...

Страница 9: ...that the patient is sitting up Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of slings should use the Amico GoLift Portable Important Pla...

Страница 10: ...arm and place the end cap back on the arm 5 Fasten all end stops tightly using an adjustable wrench NOTE If you are installing the Amico GoLift Portable in an existing track system you must ensure th...

Страница 11: ...ssible to hook on 4 Slide the top hook of the Reacher Arm through the trolley Make sure there is enough slack on the lift strap to accomplish this step otherwise it will not be possible to hook on 5 T...

Страница 12: ...ake it will override the lift and they will remain paired 2 Press and hold both the buttons on the portable lift for 5 seconds until the lift blinks Green This indicates that the lift is ready to be p...

Страница 13: ...connected Not paired Slow flashing Green N A Not connected Paired with main lift Fast flashing Green N A Maintenance Alarm At 1001 lifts the LED will flash Green To reset press Up button then Down but...

Страница 14: ...em of the lift malfunctions and the emergency lowering function is used the lift must be reset by a qualified lift technician before re use Manual Emergency Lowering The manual emergency and raising s...

Страница 15: ...of the tracks every 4 months using a damp cloth to ensure a better trolling surface for the trolley Troubleshooting Should problems arise with the use of the Amico GoLift Portable review the followin...

Страница 16: ...ual noises Inspect the hook components for any signs of damage Inspect the gears for any broken or worn teeth Verify the overspeed cam is operating freely Check emergency stop tab Check emergency lowe...

Страница 17: ...Custom Go Track Go Limb Go Turn Go GoLong Polyester SLG GO LONG WV 42X78 SLG GO LONG WV 56X78 GoLong Mesh SLG GO LONG MSH 42X78 SLG GO LONG MSH 56X78 Basic Go Long Go Active Go Comfort Go Hygiene Go C...

Страница 18: ...AT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF AMICO CLINICAL ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED DATE DATE CHECKED BY DRAWN BY NTS SCALE 2 FINISH MATERIAL YYMMDD XXX UNLESS SPECIFIED A...

Страница 19: ...ed that the lift we brought back to an area kept at room temperature between uses For example after transport in a vehicle do not leave the lift in the trunk or backseat Also try to keep the lift from...

Страница 20: ...ishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000...

Страница 21: ...IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips shor...

Страница 22: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Страница 23: ...12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimat...

Страница 24: ...months from the time of installation the warranty for power supply is one 1 year from the time of installation This warranty is valid only when the product has been properly installed as outlined in...

Страница 25: ...Notes...

Страница 26: ...Notes...

Страница 27: ...Notes...

Страница 28: ...st Beaver Creek Road Richmond Hill ON L4B 1G6 Canada 600 Prime Place Hauppauge NY 11788 USA Toll Free Phone 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email amo serv...

Отзывы: